Interferences and Limitations in Blood Glucose Self-Testing: An Overview of the Current Knowledge

Overview
Specialty Endocrinology
Date 2016 Apr 6
PMID 27044519
Citations 43
Authors
Affiliations
Soon will be listed here.
Abstract

In general, patients with diabetes performing self-monitoring of blood glucose (SMBG) can strongly rely on the accuracy of measurement results. However, various factors such as application errors, extreme environmental conditions, extreme hematocrit values, or medication interferences may potentially falsify blood glucose readings. Incorrect blood glucose readings may lead to treatment errors, for example, incorrect insulin dosing. Therefore, the diabetes team as well as the patients should be well informed about limitations in blood glucose testing. The aim of this publication is to review the current knowledge on limitations and interferences in blood glucose testing with the perspective of their clinical relevance.

Citing Articles

Development of Single-Walled Carbon Nanotube-Based Electrodes with Enhanced Dispersion and Electrochemical Properties for Blood Glucose Monitoring.

Kim D, Sobhan A, Oh J, Lee J, Park C, Lee J Biosensors (Basel). 2024; 14(12).

PMID: 39727895 PMC: 11674517. DOI: 10.3390/bios14120630.


Monitoring of the Analytical Performance of Four Different Blood Glucose Monitoring Systems: A Post-market Performance Follow-Up Study.

Mader J, Baumstark A, Tuting J, Sokol G, Schuebel R, Tong Y Diabetes Ther. 2024; 15(12):2525-2535.

PMID: 39466547 PMC: 11561228. DOI: 10.1007/s13300-024-01665-9.


Update on Patient Self-Testing with Portable and Wearable Devices: Advantages and Limitations.

Lippi G, Pighi L, Mattiuzzi C Diagnostics (Basel). 2024; 14(18).

PMID: 39335715 PMC: 11431615. DOI: 10.3390/diagnostics14182037.


Self-Powered Microfluidics for Point-of-Care Solutions: From Sampling to Detection of Proteins and Nucleic Acids.

Vloemans D, Van Hileghem L, Ordutowski H, Dal Dosso F, Spasic D, Lammertyn J Methods Mol Biol. 2024; 2804:3-50.

PMID: 38753138 DOI: 10.1007/978-1-0716-3850-7_1.


Redox-Modified Nanostructured Electrochemical Surfaces for Continuous Glucose Monitoring in Complex Biological Fluids.

Janfaza S, Radha Shanmugam N, Jolly P, Kovur P, Singh U, MacKay S Nanomaterials (Basel). 2024; 14(9).

PMID: 38727390 PMC: 11085330. DOI: 10.3390/nano14090796.


References
1.
Moore K, Vizzard N, Coleman C, McMahon J, Hayes R, Thompson C . Extreme altitude mountaineering and Type 1 diabetes; the Diabetes Federation of Ireland Kilimanjaro Expedition. Diabet Med. 2001; 18(9):749-55. DOI: 10.1046/j.0742-3071.2001.00568.x. View

2.
Bergenstal R, Ahmann A, Bailey T, Beck R, Bissen J, Buckingham B . Recommendations for standardizing glucose reporting and analysis to optimize clinical decision making in diabetes: the ambulatory glucose profile. J Diabetes Sci Technol. 2013; 7(2):562-78. PMC: 3737658. DOI: 10.1177/193229681300700234. View

3.
Louie R, Sumner S, Belcher S, Mathew R, Tran N, Kost G . Thermal stress and point-of-care testing performance: suitability of glucose test strips and blood gas cartridges for disaster response. Disaster Med Public Health Prep. 2009; 3(1):13-7. PMC: 2717087. DOI: 10.1097/DMP.0b013e3181979a06. View

4.
Baumstark A, Schmid C, Pleus S, Haug C, Freckmann G . Influence of partial pressure of oxygen in blood samples on measurement performance in glucose-oxidase-based systems for self-monitoring of blood glucose. J Diabetes Sci Technol. 2013; 7(6):1513-21. PMC: 3876329. DOI: 10.1177/193229681300700611. View

5.
Budiman E, Samant N, Resch A . Clinical implications and economic impact of accuracy differences among commercially available blood glucose monitoring systems. J Diabetes Sci Technol. 2013; 7(2):365-80. PMC: 3737638. DOI: 10.1177/193229681300700213. View