» Articles » PMID: 27013227

Dual-Quantum-Dots-Labeled Lateral Flow Strip Rapidly Quantifies Procalcitonin and C-reactive Protein

Overview
Publisher Springer
Specialty Biotechnology
Date 2016 Mar 26
PMID 27013227
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

In the article, a dual-quantum-dots-labeled (dual-QDs-labeled) lateral flow strip (LFS) method was developed for the simultaneous and rapid quantitative detection of procalcitonin (PCT) and C-reactive protein (CRP) in the blood. Two QD-antibody conjugates with different fluorescence emission spectra were produced and sprayed on the LFS to capture PCT and CRP in the blood. Furthermore, a double antibody sandwich method for PCT and, meanwhile, a competitive inhibition method for CRP were employed in the LFS. For PCT and CRP in serum assayed by the dual-QDs-labeled LFS, their detection sensitivities reached 0.1 and 1 ng/mL, respectively, and their linear quantitative detection ranges were from 0.3 to 200 ng/mL and from 50 to 250 μg/mL, respectively. There was little evidence that the PCT and CRP assays would be interfered with each other. The correlations for testing CRP and PCT in clinical samples were 99.75 and 97.02 %, respectively, between the dual-QDs-labeled LFS we developed and commercial methods. The rapid quantification of PCT and CRP on dual-QDs-labeled LFS is of great clinical value to distinguish inflammation, bacterial infection, or viral infection and to provide guidance for the use of antibiotics or other medicines.

Citing Articles

Recent Advancements in Technologies to Detect Enterohaemorrhagic Shiga Toxins.

Kim J, Lee J, Park J, Koo C, Lee M J Microbiol Biotechnol. 2023; 33(5):559-573.

PMID: 36859335 PMC: 10236172. DOI: 10.4014/jmb.2212.12025.


Simultaneous Detection of Inflammatory Biomarkers by SERS Nanotag-Based Lateral Flow Assay with Portable Cloud Raman Spectrometer.

Li Y, Liu X, Guo J, Zhang Y, Guo J, Wu X Nanomaterials (Basel). 2021; 11(6).

PMID: 34198765 PMC: 8226521. DOI: 10.3390/nano11061496.


Fluorescent hollow ZrO@CdTe nanoparticles-based lateral flow assay for simultaneous detection of C-reactive protein and troponin T.

Liu X, Ren X, Chen L, Zou J, Li T, Tan L Mikrochim Acta. 2021; 188(6):209.

PMID: 34047819 DOI: 10.1007/s00604-021-04865-3.


Duplex Shiny app quantification of the sepsis biomarkers C-reactive protein and interleukin-6 in a fast quantum dot labeled lateral flow assay.

Ruppert C, Kaiser L, Jacob L, Laufer S, Kohl M, Deigner H J Nanobiotechnology. 2020; 18(1):130.

PMID: 32912236 PMC: 7481553. DOI: 10.1186/s12951-020-00688-1.


The challenge of recognising sepsis: Future nanotechnology solutions.

Claxton A, Papafilippou L, Hadjidemetriou M, Kostarelos K, Dark P J Intensive Care Soc. 2020; 21(3):241-246.

PMID: 32782464 PMC: 7401438. DOI: 10.1177/1751143719896554.


References
1.
Schwenger V, Sis J, Breitbart A, Andrassy K . CRP levels in autoimmune disease can be specified by measurement of procalcitonin. Infection. 1998; 26(5):274-6. DOI: 10.1007/BF02962246. View

2.
Murphy C . Optical sensing with quantum dots. Anal Chem. 2002; 74(19):520A-526A. DOI: 10.1021/ac022124v. View

3.
Garnacho-Montero J, Huici-Moreno M, Gutierrez-Pizarraya A, Lopez I, Marquez-Vacaro J, Macher H . Prognostic and diagnostic value of eosinopenia, C-reactive protein, procalcitonin, and circulating cell-free DNA in critically ill patients admitted with suspicion of sepsis. Crit Care. 2014; 18(3):R116. PMC: 4229882. DOI: 10.1186/cc13908. View

4.
Chan W, Nie S . Quantum dot bioconjugates for ultrasensitive nonisotopic detection. Science. 1998; 281(5385):2016-8. DOI: 10.1126/science.281.5385.2016. View

5.
Morgenthaler N, Struck J, Fischer-Schulz C, Beier W, Bergmann A . Detection of procalcitonin (PCT) in healthy controls and patients with local infection by a sensitive ILMA. Clin Lab. 2002; 48(5-6):263-70. View