Monitoring of CRT-D Devices During Radiation Therapy in Vitro

Overview
Publisher Biomed Central
Date 2016 Mar 11
PMID 26960554
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Using of active cardiac medical devices increases steadily. In Europe, there were 183 implants of ICD and 944 implants of PM, 119 of biventricular ICD and 41 of biventricular PM, all per million inhabitants in 2014. Healthcare environments, including radiotherapy treatment rooms, are considered challenging for these implantable devices. Exposure to radiation may cause the device to experience premature elective replacement indicator, decreased pacing amplitude or pacing inhibition, inappropriate shocks or inhibition of tachyarrhythmia therapy and loss of device function. These impacts may be temporary or permanent. The aim of this study was to evaluate the influence of linear accelerator ionizing radiation dose of 10 Gy on the activity of the biventricular cardioverter-defibrillator in different position in radiation beam.

Methods: Two identical wireless communication devices with all three leads were used for the measurement. Both systems were soused into solution saline and exposed in different position in the beam of linear accelerator per 10 Gy fractions. In comparison of usually used maximum recommended dose of 2 Gy, the radiation doses used in test were five times higher. Using the simultaneous monitoring wireless communication between device and its programmer allowed watching of the devices activities, noise occurrence or drop of biventricular pacing on the programmer screen, observed by local television loop camera.

Results: At any device position in radiation beam, there were no influences of the device activity at dose of 10 Gy neither a significant increase of a solution saline temperature in any of the measured positions of CRT-D systems in linear accelerator.

Conclusions: The results of the study indicated, that the recommendation dose for treating the patients with implantable devices are too conservative and the risk of device failure is not so high. The systems can easily withstand the dose fractions of tens Gy, which would allow current single-dose-procedure treatment in radiation therapy. Even though the process of the random alteration of device memory and electrical components by scatter particles not allowed to specify a safe dose during ionizing radiation, this study showed that the safe limit are above the today used dose fractions.

Citing Articles

Safety Review of Radiotherapy for Tumor Patients with Implantable Cardiac Pacemaker.

Qiao H, Jiang F, Ning P, Zhao H, Zhao J, Zhang J Acta Cardiol Sin. 2023; 39(6):807-816.

PMID: 38022419 PMC: 10646601. DOI: 10.6515/ACS.202311_39(6).20230828A.


Safety verification of carbon-ion radiotherapy for patients with cardiac implantable electronic devices (CIEDs).

Okano N, Sakai M, Shibuya K, Tsuda K, Kanzaki T, Sano M J Radiat Res. 2021; 63(1):122-127.

PMID: 34747483 PMC: 8776694. DOI: 10.1093/jrr/rrab105.


The effect of stereotactic body radiotherapy (SBRT) using flattening filter-free beams on cardiac implantable electronic devices (CIEDs) in clinical situations.

Aslian H, Kron T, Watts T, Akalanli C, Hardcastle N, Lonski P J Appl Clin Med Phys. 2020; 21(6):121-131.

PMID: 32277741 PMC: 7324704. DOI: 10.1002/acm2.12873.


Breast Cancer Detection Using Infrared Thermal Imaging and a Deep Learning Model.

Mambou S, Maresova P, Krejcar O, Selamat A, Kuca K Sensors (Basel). 2018; 18(9).

PMID: 30149621 PMC: 6164870. DOI: 10.3390/s18092799.


Anaesthetic consideration in patients with cardiac implantable electronic devices scheduled for surgery.

Chakravarthy M, Prabhakumar D, George A Indian J Anaesth. 2017; 61(9):736-743.

PMID: 28970632 PMC: 5613599. DOI: 10.4103/ija.IJA_346_17.

References
1.
Ector H, Rickards A, Kappenberger L, Linde C, Vardas P, Oto A . The World Survey of Cardiac Pacing and Implantable Cardioverter Defibrillators: calendar year 1997--Europe. Pacing Clin Electrophysiol. 2001; 24(5):863-8. DOI: 10.1046/j.1460-9592.2001.00863.x. View

2.
Gelblum D, Amols H . Implanted cardiac defibrillator care in radiation oncology patient population. Int J Radiat Oncol Biol Phys. 2008; 73(5):1525-31. DOI: 10.1016/j.ijrobp.2008.06.1903. View

3.
Uiterwaal G, Springorum B, Scheepers E, de Ruiter G, Hurkmans C . Interference detection in implantable defibrillators induced by therapeutic radiation therapy. Neth Heart J. 2015; 14(10):330-334. PMC: 2557320. View

4.
Kusumoto F, Goldschlager N . Cardiac pacing. N Engl J Med. 1996; 334(2):89-97. DOI: 10.1056/NEJM199601113340206. View

5.
Venselaar J . The effects of ionizing radiation on eight cardiac pacemakers and the influence of electromagnetic interference from two linear accelerators. Radiother Oncol. 1985; 3(1):81-7. DOI: 10.1016/s0167-8140(85)80011-6. View