Statement Of Problem: Information is lacking about the fatigue resistance of computer-aided design and computer-aided manufacturing (CAD-CAM) tooth-colored implant custom abutment materials.

Purpose: The purpose of this in vitro study was to investigate the reliability of different types of CAD-CAM tooth-colored implant custom abutments.

Material And Methods: Zirconia (Lava Plus), lithium disilicate (IPS e.max CAD), and resin-based composite (Lava Ultimate) abutments were fabricated using CAD-CAM technology and bonded to machined titanium-6 aluminum-4 vanadium (Ti-6Al-4V) alloy inserts for conical connection implants (NobelReplace Conical Connection RP 4.3×10 mm; Nobel Biocare). Three groups (n=19) were assessed: group ZR, CAD-CAM zirconia/Ti-6Al-4V bonded abutments; group RC, CAD-CAM resin-based composite/Ti-6Al-4V bonded abutments; and group LD, CAD-CAM lithium disilicate/Ti-6Al-4V bonded abutments. Fifty-seven implant abutments were secured to implants and embedded in autopolymerizing acrylic resin according to ISO standard 14801. Static failure load (n=5) and fatigue failure load (n=14) were tested. Weibull cumulative damage analysis was used to calculate step-stress reliability at 150-N and 200-N loads with 2-sided 90% confidence limits. Representative fractured specimens were examined using stereomicroscopy and scanning electron microscopy to observe fracture patterns.

Results: Weibull plots revealed β values of 2.59 for group ZR, 0.30 for group RC, and 0.58 for group LD, indicating a wear-out or cumulative fatigue pattern for group ZR and load as the failure accelerating factor for groups RC and LD. Fractographic observation disclosed that failures initiated in the interproximal area where the lingual tensile stresses meet the compressive facial stresses for the early failure specimens. Plastic deformation of titanium inserts with fracture was observed for zirconia abutments in fatigue resistance testing.

Conclusions: Significantly higher reliability was found in group ZR, and no significant differences in reliability were determined between groups RC and LD. Differences were found in the failure characteristics of group ZR between static and fatigue loading.

Citing Articles

In Vitro Assessment of a New Block Design for Implant Crowns with Functional Gradient Fabricated with Resin Composite and Zirconia Insert.

Gutierrez Robledo N, Punset Fuste M, Rodriguez-Contreras A, Garcia Marro F, Manero Planella J, Figueras-Alvarez O Materials (Basel). 2024; 17(15).

PMID: 39124479 PMC: 11313615. DOI: 10.3390/ma17153815.


Hybrid-abutment-restoration: effect of material type on torque maintenance and fracture resistance after thermal aging.

Al-Zordk W, Elmisery A, Ghazy M Int J Implant Dent. 2020; 6(1):24.

PMID: 32578068 PMC: 7311558. DOI: 10.1186/s40729-020-00220-y.


Tensile Bond Strength of Three Custom-made Tooth-Colored Implant Superstructures to Titanium Inserts.

Zirak M, Vojdani M, Khaledi A, Farzin M J Dent (Shiraz). 2019; 20(2):131-136.

PMID: 31214642 PMC: 6538892.


A Comparative Study on the Mechanical Properties of a Polymer-Infiltrated Ceramic-Network Material Used for the Fabrication of Hybrid Abutment.

Ongun S, Kurtulmus-Yilmaz S, Meric G, Ulusoy M Materials (Basel). 2018; 11(9).

PMID: 30208618 PMC: 6164940. DOI: 10.3390/ma11091681.


Biocompatibility and antibacterial properties of zirconium nitride coating on titanium abutments: An in vitro study.

Brunello G, Brun P, Gardin C, Ferroni L, Bressan E, Meneghello R PLoS One. 2018; 13(6):e0199591.

PMID: 29944716 PMC: 6019405. DOI: 10.1371/journal.pone.0199591.