Abstract

Nucleic acid amplification-based diagnostics offer rapid, sensitive, and specific means for detecting and monitoring the progression of infectious diseases. However, this method typically requires extensive sample preparation, expensive instruments, and trained personnel. All of which hinder its use in resource-limited settings, where many infectious diseases are endemic. Here, we report on a simple, inexpensive, minimally-instrumented, smart cup platform for rapid, quantitative molecular diagnostics of pathogens at the point of care. Our smart cup takes advantage of water-triggered, exothermic chemical reaction to supply heat for the nucleic acid-based, isothermal amplification. The amplification temperature is regulated with a phase-change material (PCM). The PCM maintains the amplification reactor at a constant temperature, typically, 60-65°C, when ambient temperatures range from 12 to 35°C. To eliminate the need for an optical detector and minimize cost, we use the smartphone's flashlight to excite the fluorescent dye and the phone camera to record real-time fluorescence emission during the amplification process. The smartphone can concurrently monitor multiple amplification reactors and analyze the recorded data. Our smart cup's utility was demonstrated by amplifying and quantifying herpes simplex virus type 2 (HSV-2) with LAMP assay in our custom-made microfluidic diagnostic chip. We have consistently detected as few as 100 copies of HSV-2 viral DNA per sample. Our system does not require any lab facilities and is suitable for use at home, in the field, and in the clinic, as well as in resource-poor settings, where access to sophisticated laboratories is impractical, unaffordable, or nonexistent.

Citing Articles

Quencher-free CRISPR-based molecular detection using an amphiphilic DNA fluorescence probe.

Yang R, Guan X, Zhang J, Moon J, Guo C, Jia Z Biosens Bioelectron. 2024; 271:117054.

PMID: 39673958 PMC: 11700774. DOI: 10.1016/j.bios.2024.117054.


Compact multiplex PCR device for HIV-1 and HIV-2 viral load determination from finger-prick whole blood in resource-limited settings.

Liu T, Politza A, Ahamed M, Kshirsagar A, Zhu Y, Guan W Biosens Bioelectron. 2024; 271:116997.

PMID: 39637742 PMC: 11701890. DOI: 10.1016/j.bios.2024.116997.


Chemical Heating for Minimally Instrumented Point-of-Care (POC) Molecular Diagnostics.

Mauk M, Ansah F, El-Tholoth M Biosensors (Basel). 2024; 14(11).

PMID: 39590013 PMC: 11592221. DOI: 10.3390/bios14110554.


Enhancement of loop-mediated isothermal amplification (LAMP) with guanidine hydrochloride for the detection of subspecies (Strangles).

Knox A, Beddoe T PeerJ. 2024; 12:e17955.

PMID: 39421427 PMC: 11484460. DOI: 10.7717/peerj.17955.


Molecular Detection of Respiratory Tract Viruses in Chickens at the Point of Need by Loop-Mediated Isothermal Amplification (LAMP).

El-Tholoth M, Bau H Viruses. 2024; 16(8).

PMID: 39205222 PMC: 11359210. DOI: 10.3390/v16081248.


References
1.
Hatano B, Maki T, Obara T, Fukumoto H, Hagisawa K, Matsushita Y . LAMP using a disposable pocket warmer for anthrax detection, a highly mobile and reliable method for anti-bioterrorism. Jpn J Infect Dis. 2010; 63(1):36-40. View

2.
Lizardi P, Huang X, Zhu Z, Thomas D, Ward D . Mutation detection and single-molecule counting using isothermal rolling-circle amplification. Nat Genet. 1998; 19(3):225-32. DOI: 10.1038/898. View

3.
Xu F, Sternberg M, Kottiri B, McQuillan G, Lee F, Nahmias A . Trends in herpes simplex virus type 1 and type 2 seroprevalence in the United States. JAMA. 2006; 296(8):964-73. DOI: 10.1001/jama.296.8.964. View

4.
Laderman E, Whitworth E, Dumaual E, Jones M, Hudak A, Hogrefe W . Rapid, sensitive, and specific lateral-flow immunochromatographic point-of-care device for detection of herpes simplex virus type 2-specific immunoglobulin G antibodies in serum and whole blood. Clin Vaccine Immunol. 2007; 15(1):159-63. PMC: 2223856. DOI: 10.1128/CVI.00218-07. View

5.