Right Bundle Branch Block-Like Pattern During Uncomplicated Right Ventricular Pacing and the Effect of Pacing Site

Overview
Journal Am J Cardiol
Date 2016 Jan 23
PMID 26796192
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

Right bundle branch block (RBBB) configuration is an unexpected finding during right ventricular (RV) pacing that raises the suspicion of inadvertent left ventricular lead positioning. The aim of this study was to evaluate the prevalence of paced RBBB pattern in relation to RV lead location. This is a secondary analysis of a prospective, multicenter study, which randomized implantable cardioverter defibrillator recipients to an apical versus midseptal defibrillator lead positioning. A 12-lead electrocardiogram was recorded during intrinsic rhythm and RV pacing. Paced RBBB-like pattern was defined as positive (>0.05 mV) net amplitude of QRS complex in leads V1 and/or V2. In total, 226 patients (65.6 ± 12.0 years, 20.8% women, 53.1% apical site) were included in the study. The prevalence of paced RBBB pattern in the total population was 15.5%. A significantly lower percentage of patients in the midseptal group demonstrated RBBB-type configuration during RV pacing compared with the apical group (1.9% vs 27.5%, p <0.001). Baseline RBBB, prolonged QRS duration during intrinsic rhythm, and reduced ejection fraction were not associated with increased likelihood of paced RBBB. In the subgroup of patients with RBBB type during pacing, 91.4% of patients had a paced QRS axis from -30° to -90°, whereas 100% of patients displayed a negative QRS vector at lead V3. In conclusion, RBBB configuration is encountered in a considerable percentage of device recipients during uncomplicated RV pacing. Midseptal lead positioning is associated with significantly lower likelihood of paced RBBB pattern compared with apical location.

Citing Articles

Left ventricular septal pacing - can we trust the ECG?.

Curila K, Burri H Indian Pacing Electrophysiol J. 2023; 23(5):155-157.

PMID: 37429526 PMC: 10491966. DOI: 10.1016/j.ipej.2023.07.001.


Percutaneous extraction of a malpositioned subclavian arterial pacing lead using the retained wire technique and a vascular closure device: a case report.

Ng P, Paul V, Shetty S, Lambert J Eur Heart J Case Rep. 2022; 6(6):ytac234.

PMID: 35775017 PMC: 9240413. DOI: 10.1093/ehjcr/ytac234.


Inadvertent Lead Malposition: Prevention Is Better Than Cure.

Foley N, Sukmawati I, Harding S JACC Case Rep. 2021; 3(16):1753-1755.

PMID: 34825203 PMC: 8603047. DOI: 10.1016/j.jaccas.2021.09.008.


Prognostic impact of right bundle branch block in hospitalized patients with idiopathic dilated cardiomyopathy: a single-center cohort study.

Lai L, Jiang R, Fang W, Yan C, Tang Y, Hua W J Int Med Res. 2018; 48(1):300060518801478.

PMID: 30318986 PMC: 7287200. DOI: 10.1177/0300060518801478.


Right bundle branch block pattern after uncomplicated right ventricular outflow tract pacing in a patient with a left sided superior vena cava and corrected tetralogy of Fallot.

Gupta A, Parakh N, Juneja R Indian Pacing Electrophysiol J. 2017; 18(1):39-41.

PMID: 29183713 PMC: 5840853. DOI: 10.1016/j.ipej.2017.11.005.