Application of Intelligent Optimal KV Scanning Technology (CARE KV) in Dual-source Computed Tomography (DSCT) Coronary Angiography

Overview
Specialty General Medicine
Date 2016 Jan 16
PMID 26770354
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

This study aims to evaluate the applications and values of dual-source computed tomography (DSCT) intelligent optimal kV scanning technology (CARE kV) in coronary CT angiography (CCTA). 150 patients with normal body mass index were performed DSCT coronary angiography, then randomly divided into the "Semi", 120,100 and 80 kV Group, and the 2 "on" groups, with 30 patients in each group. The first 5 groups used the reference voltage as 120 kV, and the reference current as 400 mAs, while the other group used the reference voltage as 100 kV, and the reference current as 400 mAs. The image quality, average CT value, image noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and radiation dose were comparatively analyzed among the 5 groups. The image quality scores of the five groups showed no significant difference (P > 0.05); the average CT values and image noises had significance (P < 0.05), while SNR and CNR showed no significant difference (P > 0.05). The 80 kV group showed the biggest noise, with the CT value as 700 HU, while the radiation dose was the lowest, followed by the on group. As for the patients with normal body mass index (BMI), CARE kV-"on" could obtain high-quality images and lower radiation dose for CCTA, while the operation was simple and convenient.

Citing Articles

[Using CT to Evaluate Cardiac Function].

Lee J J Korean Soc Radiol. 2024; 85(2):308-326.

PMID: 38617866 PMC: 11009136. DOI: 10.3348/jksr.2023.0140.


Optimizing image quality and minimizing radiation dose in pediatric abdominal multiphase contrast-enhanced computed tomography: a study on CARE kV and CARE Dose 4D.

Tian X, Chang Z, Dilixiati S, Haimiti Y, Wang S, Sun J Quant Imaging Med Surg. 2024; 14(2):1985-1993.

PMID: 38415123 PMC: 10895141. DOI: 10.21037/qims-23-1181.


Validation of CARE kV automated tube voltage selection for PET-CT: PET quantification and CT radiation dose reduction in phantoms.

Bebbington N, Jorgensen T, Dupont E, Micheelsen M EJNMMI Phys. 2021; 8(1):29.

PMID: 33743091 PMC: 7981373. DOI: 10.1186/s40658-021-00373-8.


Radiation doses in CT examinations from the West China Hospital, Sichuan University and setting local diagnostic references levels.

Li Z, Zhang J, Xia C, Zhao F, Zhang K, Li Y Ann Transl Med. 2020; 8(16):1010.

PMID: 32953810 PMC: 7475485. DOI: 10.21037/atm-20-5443.


Low tube voltage prospectively ECG-triggered coronary CT angiography: a systematic review of image quality and radiation dose.

Tan S, Yeong C, Aman R, Ng K, Abdul Aziz Y, Chee K Br J Radiol. 2018; 91(1088):20170874.

PMID: 29493261 PMC: 6209486. DOI: 10.1259/bjr.20170874.

References
1.
Meinel F, Canstein C, Schoepf U, Sedlmaier M, Schmidt B, Harris B . Image quality and radiation dose of low tube voltage 3rd generation dual-source coronary CT angiography in obese patients: a phantom study. Eur Radiol. 2014; 24(7):1643-50. DOI: 10.1007/s00330-014-3194-x. View

2.
Schmidkonz C, Marwan M, Klinghammer L, Mitschke M, Schuhbaeck A, Arnold M . Interobserver variability of CT angiography for evaluation of aortic annulus dimensions prior to transcatheter aortic valve implantation (TAVI). Eur J Radiol. 2014; 83(9):1672-8. DOI: 10.1016/j.ejrad.2014.06.001. View

3.
Hell M, Bittner D, Schuhbaeck A, Muschiol G, Brand M, Lell M . Prospectively ECG-triggered high-pitch coronary angiography with third-generation dual-source CT at 70 kVp tube voltage: feasibility, image quality, radiation dose, and effect of iterative reconstruction. J Cardiovasc Comput Tomogr. 2014; 8(6):418-25. DOI: 10.1016/j.jcct.2014.09.003. View

4.
Fu F, Jin H, Feng Y . A rare case of single right coronary artery with congenital absence of left coronary artery in an adult: a case report. J Cardiothorac Surg. 2015; 10:57. PMC: 4406020. DOI: 10.1186/s13019-015-0267-0. View

5.
Koplay M, Cimen D, Sivri M, Guvenc O, Arslan D, Nayman A . Truncus arteriosus: Diagnosis with dual-source computed tomography angiography and low radiation dose. World J Radiol. 2014; 6(11):886-9. PMC: 4241496. DOI: 10.4329/wjr.v6.i11.886. View