» Articles » PMID: 26523250

Clinical and Angiographic Outcomes of Aneurysms Treated with Two Self-expanding Stent-assisted Coiling Systems: A Comparison of Solitaire AB and Enterprise VRD Stents

Overview
Date 2015 Nov 3
PMID 26523250
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: The purpose of this study was to compare clinical findings and outcomes of Enterprise and Solitaire stent-assisted coiling (SAC).

Materials And Methods: Between January 2012 and March 2014, 86 patients (mean age, 60.3 years) harboring 89 aneurysms were treated with Enterprise (n = 57) or Solitaire (n = 32) SAC. The patients' demographics, angiographic results, and clinical outcomes were reviewed retrospectively.

Results: There were no cases of stent navigation, deployment failure, arterial dissection, or intraoperative aneurysmal rupture. Angiographic follow-up imaging was available for 86 (96.6%) aneurysms (Enterprise group, n = 55; Solitaire group, n = 31). Immediate postoperative and follow-up angiographic results showed no flow or only minimal flow into the neck in 83% (Enterprise group, 77.2%; Solitaire group, 93.8%) and 95.3% (Enterprise group, 92.7%; Solitaire group, 100%) of SAC-treated aneurysms, respectively. Both stent groups showed good immediate postoperative and follow-up clinical outcomes. Excepting 2 cases, all patients achieved modified Rankin Scale scores of 0. Coil loop or tail protrusion into the parent artery was observed in 17 (29.8%) and 7 (21.9%) cases in the Enterprise and Solitaire groups, respectively. No statistically significant difference in terms of angiographic results or clinical outcomes was observed between the groups.

Conclusion: Excellent and comparable clinical and angiographic outcomes for wide-neck intracranial aneurysms were achieved using both stents. Because of its higher radial strength and better vessel wall apposition, we cautiously propose that the Solitaire stent may be more effective for SAC of aneurysms harboring a large or severe tortuous parent artery.

Citing Articles

Comparative Outcome Analysis of Enterprise and Neuroform Stent-Assisted Coiling of Cerebral Aneurysms: A Review of the Literature.

Goyal R, Kato Y, Kawase T, Suzuki K, Yamada Y, Sharma S Asian J Neurosurg. 2020; 15(1):4-9.

PMID: 32181165 PMC: 7057877. DOI: 10.4103/ajns.AJNS_284_19.


Does stent type impact coil embolization outcomes in extended follow-up of small-sized aneurysms (< 10 mm)?.

Lee J, Cho Y, Yoo D, Kang H, Cho W, Kim J Neuroradiology. 2018; 60(7):747-756.

PMID: 29675589 DOI: 10.1007/s00234-018-2022-4.


Stent-assisted coiling of cerebral aneurysms with the Neuroform Atlas stent.

Cay F, Peker A, Arat A Interv Neuroradiol. 2018; 24(3):263-269.

PMID: 29350091 PMC: 5967178. DOI: 10.1177/1591019917753710.


Stent-assisted coiling of intracranial aneurysms using LEO stents: long-term follow-up in 153 patients.

Sedat J, Chau Y, Gaudart J, Sachet M, Beuil S, Lonjon M Neuroradiology. 2017; 60(2):211-219.

PMID: 29273960 DOI: 10.1007/s00234-017-1965-1.

References
1.
Klisch J, Clajus C, Sychra V, Eger C, Strasilla C, Rosahl S . Coil embolization of anterior circulation aneurysms supported by the Solitaire AB Neurovascular Remodeling Device. Neuroradiology. 2009; 52(5):349-59. DOI: 10.1007/s00234-009-0568-x. View

2.
Geyik S, Yavuz K, Yurttutan N, Saatci I, Cekirge H . Stent-assisted coiling in endovascular treatment of 500 consecutive cerebral aneurysms with long-term follow-up. AJNR Am J Neuroradiol. 2013; 34(11):2157-62. PMC: 7964833. DOI: 10.3174/ajnr.A3574. View

3.
Fargen K, Hoh B, Welch B, Pride G, Lanzino G, Boulos A . Long-term results of enterprise stent-assisted coiling of cerebral aneurysms. Neurosurgery. 2012; 71(2):239-44. DOI: 10.1227/NEU.0b013e3182571953. View

4.
Klisch J, Eger C, Sychra V, Strasilla C, Basche S, Weber J . Stent-assisted coil embolization of posterior circulation aneurysms using solitaire ab: preliminary experience. Neurosurgery. 2009; 65(2):258-66. DOI: 10.1227/01.NEU.0000348295.44970.C8. View

5.
Ebrahimi N, Claus B, Lee C, Biondi A, Benndorf G . Stent conformity in curved vascular models with simulated aneurysm necks using flat-panel CT: an in vitro study. AJNR Am J Neuroradiol. 2007; 28(5):823-9. PMC: 8134344. View