» Articles » PMID: 26472513

Do We Really Need to Wear Proper Eye Protection When Using Holmium:YAG Laser During Endourologic Procedures? Results from an Ex Vivo Animal Model on Pig Eyes

Overview
Journal J Endourol
Publisher Mary Ann Liebert
Date 2015 Oct 17
PMID 26472513
Citations 14
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: We sought to evaluate the effect of holmium:yttrium-aluminum-garnet (Ho:YAG) laser exposure on ex vivo pig eyes and to test the protective action of different glasses in preventing eye lesions in case of accident.

Materials And Methods: We pointed the tip of a Ho:YAG laser fiber from different distances (0, 3, 5, 8, 10, and 20 cm, respectively) toward the center of the pupil of the pig eye. The Ho:YAG laser was activated for 1 or 5 seconds at three different settings (0.5 J-20 Hz, 1 J-10 Hz, and 2 J-10 Hz, respectively). The experiment was repeated using laser safety glasses and eyeglasses. A total of 78 pig eyes were used. The effects of the Ho:YAG laser on pig eyes were assessed by histopathology. Comparable laser emission experiments were performed on thermal paper at different distances using different pulse energies.

Results: Ho:YAG laser-induced corneal lesions were observed in unprotected eyes, ranging from superficial burning lesions to full-thickness necrotic areas, and were directly related to pulse energy and time of exposure and inversely related to the distance from the eye. When the laser was placed 5 cm or more, no corneal damage was observed regardless of the laser setting and the time of exposure. Similar distance/energy level relationships were observed on thermal paper. No damage was observed to the lens or the retina in any of the Ho-YAG laser-treated eyes or in any of the eyes protected by laser safety and eyeglasses.

Conclusions: Ho:YAG lasers can cause damage when set to high energy, but only to the cornea, from close distances (0-5 cm) and in the absence of eye protection. Eyeglasses are equally effective in preventing laser damage as laser safety glasses.

Citing Articles

The Impact of Corneal Oedema on UV Light Transmission: An Experimental Study in Porcine Eyes.

Briceno-Lopez C, Garcia-Domene M, Peris-Martinez C, Luque-Cobija M, Diez-Ajenjo M, Burguera-Gimenez N J Clin Med. 2024; 13(23).

PMID: 39685687 PMC: 11642682. DOI: 10.3390/jcm13237228.


Preconceived Impressions Regarding Holmium:YAG Laser Safety in the Urology Operating Theatre.

Barns M, Magee D, England T Res Rep Urol. 2024; 16:143-150.

PMID: 38912094 PMC: 11192151. DOI: 10.2147/RRU.S457617.


Successful endoscopic laser lithotripsy in 2 cases of Bouveret syndrome and cholecystocolonic fistulae-induced colonic obstruction: a minimally invasive approach.

Vega E, Rivera B, Harz C VideoGIE. 2024; 9(6):290-294.

PMID: 38887731 PMC: 11180364. DOI: 10.1016/j.vgie.2024.03.001.


What to expect from the novel pulsed thulium:YAG laser? A systematic review of endourological applications.

Ventimiglia E, Robesti D, Bevilacqua L, Tondelli E, Oliva I, Orecchia L World J Urol. 2023; 41(11):3301-3308.

PMID: 37682286 DOI: 10.1007/s00345-023-04580-z.


Holmium and Thulium Fiber Laser Safety in Endourological Practice: What Does the Clinician Need to Know?.

Juliebo-Jones P, Somani B, Gjengsto P, Asoy M, Beisland C, Ulvik O Curr Urol Rep. 2023; 24(9):409-415.

PMID: 37256486 PMC: 10449946. DOI: 10.1007/s11934-023-01168-3.