Calibration and Evaluation of a Magnetically Tracked ICE Probe for Guidance of Left Atrial Ablation Therapy

Overview
Date 2015 Sep 25
PMID 26401065
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

The novel prototype system for advanced visualization for image-guided left atrial ablation therapy developed in our laboratory permits ready integration of multiple imaging modalities, surgical instrument tracking, interventional devices and electro-physiologic data. This technology allows subject-specific procedure planning and guidance using 3D dynamic, patient-specific models of the patient's heart, augmented with real-time intracardiac echocardiography (ICE). In order for the 2D ICE images to provide intuitive visualization for accurate catheter to surgical target navigation, the transducer must be tracked, so that the acquired images can be appropriately presented with respect to the patient-specific anatomy. Here we present the implementation of a previously developed ultrasound calibration technique for a magnetically tracked ICE transducer, along with a series of evaluation methods to ensure accurate imaging and faithful representation of the imaged structures. Using an engineering-designed phantom, target localization accuracy is assessed by comparing known target locations with their transformed locations inferred from the tracked US images. In addition, the 3D volume reconstruction accuracy is also estimated by comparing a truth volume to that reconstructed from sequential 2D US images. Clinically emulating validation studies are conducted using a patient-specific left atrial phantom. Target localization error of clinically-relevant surgical targets represented by nylon fiducials implanted within the endocardial wall of the phantom was assessed. Our studies have demonstrated 2.4 0.8 mm target localization error in the engineering-designed evaluation phantoms, 94.8 4.6 % volume reconstruction accuracy, and 3.1 1.2 mm target localization error in the left atrial-mimicking phantom. These results are consistent with those disseminated in the literature and also with the accuracy constraints imposed by the employed technology and the clinical application.

Citing Articles

On mixed reality environments for minimally invasive therapy guidance: systems architecture, successes and challenges in their implementation from laboratory to clinic.

Linte C, Davenport K, Cleary K, Peters C, Vosburgh K, Navab N Comput Med Imaging Graph. 2013; 37(2):83-97.

PMID: 23632059 PMC: 3796657. DOI: 10.1016/j.compmedimag.2012.12.002.

References
1.
Khamene A, Sauer F . A novel phantom-less spatial and temporal ultrasound calibration method. Med Image Comput Comput Assist Interv. 2006; 8(Pt 2):65-72. DOI: 10.1007/11566489_9. View

2.
Comeau R, Sadikot A, Fenster A, Peters T . Intraoperative ultrasound for guidance and tissue shift correction in image-guided neurosurgery. Med Phys. 2000; 27(4):787-800. DOI: 10.1118/1.598942. View

3.
Rohling R, Munger P, Hollerbach J, Peter T . Comparison of relative accuracy between a mechanical and an optical position tracker for image-guided neurosurgery. J Image Guid Surg. 1995; 1(1):30-4. DOI: 10.1002/(SICI)1522-712X(1995)1:1<30::AID-IGS5>3.0.CO;2-N. View

4.
Dong J, Dickfeld T, Lamiy S, Calkins H . Catheter ablation of atrial fibrillation guided by registered computed tomographic image of the atrium. Heart Rhythm. 2005; 2(9):1021-2. DOI: 10.1016/j.hrthm.2005.05.004. View

5.
Haissaguerre M, Jais P, Shah D, Gencel L, Pradeau V, Garrigues S . Right and left atrial radiofrequency catheter therapy of paroxysmal atrial fibrillation. J Cardiovasc Electrophysiol. 1996; 7(12):1132-44. DOI: 10.1111/j.1540-8167.1996.tb00492.x. View