An Improved Electrical Impedance Myography (EIM) Tongue Array for Use in Clinical Trials
Overview
Psychiatry
Affiliations
Objectives: Electrical impedance myography (EIM) measurements of the tongue could provide valuable information about bulbar dysfunction in amyotrophic lateral sclerosis (ALS). A prototype tongue depressor EIM array produced gag reflexes. The objectives of this study were to determine the reliability, mean phase values, and tolerability of tongue EIM measurements using a smaller electrode array.
Methods: Tongue EIM measurements were performed in a total of 31 healthy individuals and four neuromuscular patients with lingual abnormalities. Reliability was assessed by calculating the intraclass correlation coefficient (ICC) and percent difference in addition to performing Bland-Altman analyses. Standard descriptive statistics, including results of a Mann-Whitney test, were also determined.
Results: At the 50 kHz frequency, the ICCs for intra- and inter-rater reliability were 0.76 with 5.17% difference and 0.78 with 5.34% difference respectively. The mean EIM phase values of healthy participants (11.61° ± 1.00°) and patients (9.87° ± 1.28°) were significantly different (p=0.0051). None of the participants experienced gag reflexes or discomfort.
Conclusions: The small tongue array provided good inter- and intra-rater reliability, could preliminarily distinguish between healthy and diseased muscle, and was well-tolerated.
Significance: Biomarker information about tongue health could be more comfortably obtained with a smaller EIM array.
Hansen N, Woodman K, Buoy S, Mao S, Barbon C, Lai S Head Neck. 2023; 46(3):581-591.
PMID: 38133080 PMC: 10922903. DOI: 10.1002/hed.27618.
Luo X, Victoria Gutierrez Pulido H, Rutkove S, Sanchez B IEEE Open J Eng Med Biol. 2022; 2:278-285.
PMID: 35402988 PMC: 8940204. DOI: 10.1109/OJEMB.2021.3117871.
Critcher S, Freeborn T IEEE Open J Circuits Syst. 2021; 2:732-742.
PMID: 34901875 PMC: 8656411. DOI: 10.1109/OJCAS.2021.3122369.
Critcher S, Freeborn T Sensors (Basel). 2021; 21(9).
PMID: 33923037 PMC: 8123364. DOI: 10.3390/s21093013.
In vivo muscle conduction study of the tongue using a multi-electrode tongue depressor.
Luo X, Victoria Gutierrez Pulido H, Rutkove S, Sanchez B Clin Neurophysiol. 2020; 132(2):683-687.
PMID: 33309468 PMC: 7880865. DOI: 10.1016/j.clinph.2020.11.010.