Imposed Work of Breathing for Flow Meters with In-Line Versus Flow-Through Technique During Simulated Neonatal Breathing

Overview
Journal PLoS One
Date 2015 Jul 21
PMID 26192188
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The ability to determine airflow during nasal CPAP (NCPAP) treatment without adding dead space or resistance would be useful when investigating the physiologic effects of different NCPAP systems on breathing. The aim of this study was to investigate the effect on pressure stability of different flow measuring devices at the in-line and flow-through position, using simulated neonatal breathing.

Methods: Six different flow measure devices were evaluated by recording pressure changes and imposed work of breathing for breaths with 16 and 32 ml tidal volumes. The tests were performed initially with the devices in an in line position and with 5 and 10 L/min using flow through technique, without CPAP. The flow meters were then subsequently tested with an Infant Flow CPAP system at 3, 5 and 8 cm H2O pressure using flow through technique. The quality of the recorded signals was compared graphically.

Results: The resistance of the measuring devices generated pressure swings and imposed work of breathing. With bias flow, the resistance also generated CPAP pressure. Three of the devices had low resistance and generated no changes in pressure stability or CPAP pressure. The two devices intended for neonatal use had the highest measured resistance.

Conclusion: The importance of pressure stability and increased work of breathing during non-invasive respiratory support are insufficiently studied. Clinical trials using flow-through technique have not focused on pressure stability. Our results indicate that a flow-through technique might be a way forward in obtaining a sufficiently high signal quality without the added effects of rebreathing and increased work of breathing. The results should stimulate further research and the development of equipment for dynamic flow measurements in neonates.

Citing Articles

Do newborn infants exhale through the CPAP system? Secondary analysis of a randomised cross-over trial.

Gunnarsdottir K, Falk M, Baldursdottir S, Donaldsson S, Jonsson B, Drevhammar T Arch Dis Child Fetal Neonatal Ed. 2022; 108(3):232-236.

PMID: 36261145 PMC: 10176426. DOI: 10.1136/archdischild-2022-324462.


Interface leakage during neonatal CPAP treatment: a randomised, cross-over trial.

Falk M, Gunnarsdottir K, Baldursdottir S, Donaldsson S, Jonsson B, Drevhammar T Arch Dis Child Fetal Neonatal Ed. 2021; 106(6):663-667.

PMID: 33963004 PMC: 8543197. DOI: 10.1136/archdischild-2021-321579.

References
1.
Sandberg K, Hjalmarson O . CPAP of 4 cm H(2)O Has no short-term benefit at term in infants with BPD. Neonatology. 2012; 102(4):282-6. DOI: 10.1159/000341765. View

2.
Drevhammar T, Nilsson K, Zetterstrom H, Jonsson B . Comparison of seven infant continuous positive airway pressure systems using simulated neonatal breathing. Pediatr Crit Care Med. 2011; 13(2):e113-9. DOI: 10.1097/PCC.0b013e31822f1b79. View

3.
Drevhammar T, Nilsson K, Zetterstrom H, Jonsson B . Comparison of nasal continuous positive airway pressure delivered by seven ventilators using simulated neonatal breathing. Pediatr Crit Care Med. 2013; 14(4):e196-201. DOI: 10.1097/PCC.0b013e31827212e4. View

4.
Fouzas S, Hacki C, Latzin P, Proietti E, Schulzke S, Frey U . Volumetric capnography in infants with bronchopulmonary dysplasia. J Pediatr. 2013; 164(2):283-8.e1-3. DOI: 10.1016/j.jpeds.2013.09.034. View

5.
Foitzik B, Schmidt M, Proquitte H, Schmalisch G . Deadspace free ventilatory measurements in newborns during mechanical ventilation. Crit Care Med. 2001; 29(2):413-9. DOI: 10.1097/00003246-200102000-00036. View