Inter- and Intrafraction Target Motion in Highly Focused Single Vocal Cord Irradiation of T1a Larynx Cancer Patients

Overview
Specialties Oncology
Radiology
Date 2015 Jun 22
PMID 26094125
Citations 13
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The purpose of this study was to verify clinical target volume--planning target volume (CTV-PTV) margins in single vocal cord irradiation (SVCI) of T1a larynx tumors and characterize inter- and intrafraction target motion.

Methods And Materials: For 42 patients, a single vocal cord was irradiated using intensity modulated radiation therapy at a total dose of 58.1 Gy (16 fractions × 3.63 Gy). A daily cone beam computed tomography (CBCT) scan was performed to online correct the setup of the thyroid cartilage after patient positioning with in-room lasers (interfraction motion correction). To monitor intrafraction motion, CBCT scans were also acquired just after patient repositioning and after dose delivery. A mixed online-offline setup correction protocol ("O2 protocol") was designed to compensate for both inter- and intrafraction motion.

Results: Observed interfraction, systematic (Σ), and random (σ) setup errors in left-right (LR), craniocaudal (CC), and anteroposterior (AP) directions were 0.9, 2.0, and 1.1 mm and 1.0, 1.6, and 1.0 mm, respectively. After correction of these errors, the following intrafraction movements derived from the CBCT acquired after dose delivery were: Σ = 0.4, 1.3, and 0.7 mm, and σ = 0.8, 1.4, and 0.8 mm. More than half of the patients showed a systematic non-zero intrafraction shift in target position, (ie, the mean intrafraction displacement over the treatment fractions was statistically significantly different from zero; P<.05). With the applied CTV-PTV margins (for most patients 3, 5, and 3 mm in LR, CC, and AP directions, respectively), the minimum CTV dose, estimated from the target displacements observed in the last CBCT, was at least 94% of the prescribed dose for all patients and more than 98% for most patients (37 of 42). The proposed O2 protocol could effectively reduce the systematic intrafraction errors observed after dose delivery to almost zero (Σ = 0.1, 0.2, 0.2 mm).

Conclusions: With adequate image guidance and CTV-PTV margins in LR, CC, and AP directions of 3, 5, and 3 mm, respectively, excellent target coverage in SVCI could be ensured.

Citing Articles

Pencil Beam Scanning Proton Therapy as Single Vocal Cord Irradiation for Early-Stage Glottic Cancer.

Savla B, Jatczak J, Molitoris J, Witek M, Marter K, Zakhary M Int J Part Ther. 2024; 13:100623.

PMID: 39139470 PMC: 11321307. DOI: 10.1016/j.ijpt.2024.100623.


Assessment and validation of glottic motion using cone-beam CT and real-time cine MRI.

Chun S, Son J, Kang S, Choi C, Kim J, Kim Y Strahlenther Onkol. 2024; 200(5):418-424.

PMID: 38488899 PMC: 11039518. DOI: 10.1007/s00066-024-02204-y.


Real-Time Tracking of Laryngeal Motion via the Surface Depth-Sensing Technique for Radiotherapy in Laryngeal Cancer Patients.

Lee W, Leu Y, Chen J, Dai K, Hou T, Chang C Bioengineering (Basel). 2023; 10(8).

PMID: 37627793 PMC: 10451758. DOI: 10.3390/bioengineering10080908.


VoiceS: voice quality after transoral CO laser surgery versus single vocal cord irradiation for unilateral stage 0 and I glottic larynx cancer-a randomized phase III trial.

Reinhardt P, Giger R, Seifert E, Shelan M, Riggenbach E, Terribilini D Trials. 2022; 23(1):906.

PMID: 36303192 PMC: 9615245. DOI: 10.1186/s13063-022-06841-5.


Early-Stage Glottic Squamous Cell Carcinoma in the Era of Image-Guided Radiotherapy.

Gupta A, Wong K, Newbold K, Bhide S, Nutting C, Harrington K Front Oncol. 2021; 11:753908.

PMID: 34616688 PMC: 8488425. DOI: 10.3389/fonc.2021.753908.