Investigation of Collimator Influential Parameter on SPECT Image Quality: a Monte Carlo Study

Overview
Date 2015 May 15
PMID 25973410
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Obtaining high quality images in Single Photon Emission Tomography (SPECT) device  is the most important goal in nuclear medicine. Because if image quality is low, the possibility of making a mistake in diagnosing and treating the patient will rise. Studying effective factors in spatial resolution of imaging systems is thus deemed to be vital. One of the most important factors in SPECT imaging in nuclear medicine is the use of an appropriate collimator for a certain radiopharmaceutical feature in order to create the best image as it can be effective in the quantity of Full Width at Half Maximum (FWHM) which is the main parameter  in spatial resolution.

Method: In this research, the simulation of the detector and collimator of SPECT imaging device, Model HD3 made by Philips Co. and the investigation of important factors on the collimator were carried out using MCNP-4c code.

Results: The results of the experimental measurments and simulation calculations revealed a relative difference of less than 5% leading to the confirmation of the accuracy of conducted simulation MCNP code calculation.

Conclusion: This is the first essential step in the design and modelling of new collimators used for creating high quality images in nuclear medicine.

References
1.
Chun S, Fessler J, Dewaraja Y . Correction for collimator-detector response in SPECT using point spread function template. IEEE Trans Med Imaging. 2012; 32(2):295-305. PMC: 3619230. DOI: 10.1109/TMI.2012.2225441. View

2.
Maniawski P, Morgan H, Wackers F . Orbit-related variation in spatial resolution as a source of artifactual defects in thallium-201 SPECT. J Nucl Med. 1991; 32(5):871-5. View

3.
Seo Y, Wong K, Sun M, Franc B, Hawkins R, Hasegawa B . Correction of photon attenuation and collimator response for a body-contouring SPECT/CT imaging system. J Nucl Med. 2005; 46(5):868-77. View

4.
Glick S, Penney B, King M, Byrne C . Noniterative compensation for the distance-dependent detector response and photon attenuation in SPECT imaging. IEEE Trans Med Imaging. 1994; 13(2):363-74. DOI: 10.1109/42.293929. View

5.
Hutton B, Lau Y . Application of distance-dependent resolution compensation and post-reconstruction filtering for myocardial SPECT. Phys Med Biol. 1998; 43(6):1679-93. DOI: 10.1088/0031-9155/43/6/022. View