Effect of Reconstruction Algorithms on the Accuracy of (99m)Tc Sestamibi SPECT/CT Parathyroid Imaging

Overview
Date 2015 May 15
PMID 25973340
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

The superiority of SPECT/CT over SPECT for (99m)Tc-sestamibi parathyroid imaging often is assumed to be due to improved lesion localization provided by the anatomic component (computed tomography) of the examination. It also is possible that this superiority may be related to the algorithms used for SPECT data reconstruction. The objective of this investigation was to determine the effect of SPECT reconstruction algorithms on the accuracy of MIBI SPECT/CT parathyroid imaging. We retrospectively analyzed preoperative MIBI SPECT/CT parathyroid imaging studies performed on 106 patients. SPECT data were reconstructed by filtered back projection (FBP) and by iterative reconstruction with corrections for collimator resolution recovery and attenuation (IRC). Two experienced readers independently graded lesion detection certainty on a 5-point scale without knowledge of each other's readings, reconstruction methods, other test results or final diagnoses. All patients had surgical confirmation of the final diagnosis, including disease limited to the neck, and location and weight of excised lesion(s). There were 135 parathyroid lesions among the 106 patients. For FBP SPECT/CT and IRC SPECT/CT sensitivity was 76% and 90% (p = 0.003), specificity was 87% and 87% (p = 0.90), and accuracy was 83% and 88% (p = 0.04), respectively. Inter-rater agreement was significantly higher for IRC than for FBP (kappa = 0.76, "good agreement", versus kappa = 0.58, "moderate agreement", p < 0.0001). We conclude that the improved accuracy of MIBI SPECT/CT compared to MIBI SPECT for preoperative parathyroid lesion localization is due in part to the use of IRC for SPECT data reconstruction.

Citing Articles

Preoperative injection of carbon nanoparticles is beneficial to the patients with thyroid papillary carcinoma: From a prospective study of 102 cases.

Yan S, Zhao W, Wang B, Zhang L Medicine (Baltimore). 2018; 97(27):e11364.

PMID: 29979421 PMC: 6076069. DOI: 10.1097/MD.0000000000011364.


Tc-Sestamibi/I Subtraction SPECT/CT in Parathyroid Scintigraphy: Is Additional Pinhole Imaging Useful?.

Tunninen V, Varjo P, Kauppinen T, Holm A, Eskola H, Seppanen M Int J Mol Imaging. 2017; 2017:2712018.

PMID: 29181196 PMC: 5664318. DOI: 10.1155/2017/2712018.


Intraoperative localization of the parathyroid glands with indocyanine green and Firefly(R) technology during BABA robotic thyroidectomy.

Yu H, Chung J, Yi J, Song R, Lee J, Kwon H Surg Endosc. 2016; 31(7):3020-3027.

PMID: 27864717 DOI: 10.1007/s00464-016-5330-y.

References
1.
Daou D, Pointurier I, Coaguila C, Vilain D, Benada A, Lebtahi R . Performance of OSEM and depth-dependent resolution recovery algorithms for the evaluation of global left ventricular function in 201Tl gated myocardial perfusion SPECT. J Nucl Med. 2003; 44(2):155-62. View

2.
Nichols K, Tronco G, Tomas M, Kunjummen B, Siripun L, Rini J . Phantom experiments to improve parathyroid lesion detection. Med Phys. 2008; 34(12):4792-7. DOI: 10.1118/1.2804553. View

3.
Prommegger R, Wimmer G, Profanter C, Sauper T, Sieb M, Kovacs P . Virtual neck exploration: a new method for localizing abnormal parathyroid glands. Ann Surg. 2009; 250(5):761-5. DOI: 10.1097/SLA.0b013e3181bd906b. View

4.
Pan T, Luo D, Kohli V, King M . Influence of OSEM, elliptical orbits and background activity on SPECT 3D resolution recovery. Phys Med Biol. 1998; 42(12):2517-29. DOI: 10.1088/0031-9155/42/12/015. View

5.
Nichols K, Tomas M, Tronco G, Rini J, Kunjummen B, Heller K . Preoperative parathyroid scintigraphic lesion localization: accuracy of various types of readings. Radiology. 2008; 248(1):221-32. DOI: 10.1148/radiol.2481071066. View