» Articles » PMID: 25888086

Ex-vivo Perfusion of Donor Hearts for Human Heart Transplantation (PROCEED II): a Prospective, Open-label, Multicentre, Randomised Non-inferiority Trial

Overview
Journal Lancet
Publisher Elsevier
Specialty General Medicine
Date 2015 Apr 19
PMID 25888086
Citations 133
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The Organ Care System is the only clinical platform for ex-vivo perfusion of human donor hearts. The system preserves the donor heart in a warm beating state during transport from the donor hospital to the recipient hospital. We aimed to assess the clinical outcomes of the Organ Care System compared with standard cold storage of human donor hearts for transplantation.

Methods: We did this prospective, open-label, multicentre, randomised non-inferiority trial at ten heart-transplant centres in the USA and Europe. Eligible heart-transplant candidates (aged >18 years) were randomly assigned (1:1) to receive donor hearts preserved with either the Organ Care System or standard cold storage. Participants, investigators, and medical staff were not masked to group assignment. The primary endpoint was 30 day patient and graft survival, with a 10% non-inferiority margin. We did analyses in the intention-to-treat, as-treated, and per-protocol populations. This trial is registered with ClinicalTrials.gov, number NCT00855712.

Findings: Between June 29, 2010, and Sept 16, 2013, we randomly assigned 130 patients to the Organ Care System group (n=67) or the standard cold storage group (n=63). 30 day patient and graft survival rates were 94% (n=63) in the Organ Care System group and 97% (n=61) in the standard cold storage group (difference 2·8%, one-sided 95% upper confidence bound 8·8; p=0·45). Eight (13%) patients in the Organ Care System group and nine (14%) patients in the standard cold storage group had cardiac-related serious adverse events.

Interpretation: Heart transplantation using donor hearts adequately preserved with the Organ Care System or with standard cold storage yield similar short-term clinical outcomes. The metabolic assessment capability of the Organ Care System needs further study.

Funding: TransMedics.

Citing Articles

Clinical Outcomes of Machine Perfusion and Temperature Control Systems in Heart Transplantation: Where We Stand.

Nasim U, Dorken-Gallastegi A, Dadson P, Hong Y J Clin Med. 2025; 14(4).

PMID: 40004682 PMC: 11856903. DOI: 10.3390/jcm14041152.


Donor Heart Preservation for Heart Transplantation: Single-Center Experience with Three Different Techniques.

Lechiancole A, Gliozzi G, Sponga S, Visentin P, Beltrami A, Piani D J Clin Med. 2025; 14(4).

PMID: 40004640 PMC: 11856374. DOI: 10.3390/jcm14041108.


Donation after circulatory death transplantation: a systematic review and meta-analysis of outcomes and methods of donation.

Jolliffe J, Brookes J, Williams M, Walker E, Jansz P, Watson A Ann Cardiothorac Surg. 2025; 14(1):11-27.

PMID: 39944506 PMC: 11811570. DOI: 10.21037/acs-2024-dcd-0132.


Effects of Nighttime Procurement and Transplantation on Outcomes Following Heart Transplantation.

Hong Y, Hess N, Dorken-Gallastegi A, Abdullah M, Iyanna N, Nasim U Clin Transplant. 2025; 39(2):e70093.

PMID: 39887494 PMC: 11781293. DOI: 10.1111/ctr.70093.


Involving Patients in Hospital-Based Health Technology Assessment of Innovative Medical Devices: Adapting to a Specific Local Context and Lessons Learned From the Assessment of an Ex Vivo Perfusion System of Human Donor Hearts.

Atfeh J, Guerre P, Berkesse A, Thual G, Pozzi M, Huot L Health Expect. 2024; 27(6):e70119.

PMID: 39660693 PMC: 11632624. DOI: 10.1111/hex.70119.