Effect of Increased Crown Height on Stress Distribution in Short Dental Implant Components and Their Surrounding Bone: A Finite Element Analysis

Overview
Journal J Prosthet Dent
Specialty Dentistry
Date 2015 Mar 22
PMID 25794917
Citations 22
Authors
Affiliations
Soon will be listed here.
Abstract

Statement Of Problem: Implants in posterior regions of the jaw require short dental implants with long crown heights, leading to increased crown-to-implant ratios and mechanical stress. This can lead to fracture and screw loosening.

Purpose: The purpose of this study was to investigate the dynamic nature and behavior of prosthetic components and preimplant bone and evaluate the effect of increased crown height space (CHS) and crown-to-implant ratio on stress concentrations under external oblique forces.

Material And Methods: The severely resorbed bone of a posterior mandible site was modeled with Mimics and Catia software. A second mandibular premolar tooth was modeled with CHS values of 8.8, 11.2, 13.6, and 16 mm. A Straumann implant (4.1×8 mm), a directly attached crown, and an abutment screw were modeled with geometric data and designed by using SolidWorks software. Abaqus software was used for the dynamic simulation of screw tightening and the application of an external load to the buccal cusp at a 75.8-degree angle with the occlusal plane. The distribution of screw load and member load at each step was compared, and the stress values were calculated within the dental implant complex and surrounding bone.

Results: During tightening, the magnitude and distribution of the preload and clamp load were uniform and equal at the cross section of all CHSs. Under an external load, the screw load decreased and member load increased. An increase in the CHS caused the corresponding distribution to become more nonuniform and increased the maximum compressive and tensile stresses in the preimplant bone. Additionally, the von Mises stress decreased at the abutment screw and increased at the abutment and fixture.

Conclusions: Under nonaxial forces, increased CHS does not influence the decrease in screw load or increase in member load. However, it contributes to screw loosening and fatigue fracture by skewing the stress distribution to the transverse section of the implant.

Citing Articles

Investigation of the cushioning mechanism of a novel dental implant system with composite hydrogel.

Chi C, Chung T, Wu Y, Chen W, Lin C J Dent Sci. 2025; 20(1):61-68.

PMID: 39873077 PMC: 11763535. DOI: 10.1016/j.jds.2024.10.016.


Impact Testing in Implant-Supported Prostheses and Natural Teeth: A Systematic Review of Properties and Performance.

Marti-Vigil J, Casamitjana J, Marimon X, Cerrolaza M, Medina-Galvez R, Canto-Naves O Materials (Basel). 2024; 17(16).

PMID: 39203218 PMC: 11356274. DOI: 10.3390/ma17164040.


Stress Distribution on Short Implants with Varying Crown Heights - An Study.

Hariharan A, Krithika A, Thanya K, Thamarai C, Parameswari D, Dhevishri S J Pharm Bioallied Sci. 2024; 16(Suppl 2):S1498-S1504.

PMID: 38882717 PMC: 11174330. DOI: 10.4103/jpbs.jpbs_1111_23.


The effect of different implant-abutment types and heights on screw loosening in cases with increased crown height space.

Hendi A, Mirzaee S, Falahchai M Clin Exp Dent Res. 2024; 10(3):e894.

PMID: 38881218 PMC: 11180708. DOI: 10.1002/cre2.894.


Advancing 3D Dental Implant Finite Element Analysis: Incorporating Biomimetic Trabecular Bone with Varied Pore Sizes in Voronoi Lattices.

Alemayehu D, Todoh M, Huang S J Funct Biomater. 2024; 15(4).

PMID: 38667551 PMC: 11051206. DOI: 10.3390/jfb15040094.