Prevention of Vertebral Body-splitting Fractures After Multilevel ProDisc-L Implantation

Overview
Date 2015 Feb 20
PMID 25694877
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Prior studies of multilevel ProDisc-L (PD-L) implants (Synthes Spine, Inc., West Chester, Pennsylvania) using the standard US technique have used conventional radiography postoperatively. We found vertebral body-splitting fractures (VB-SFs) in interposed vertebral bodies after 5 sequential multilevel PD-L device implantations using the standard US technique. These were identified with postoperative computed tomography (CT) but were not visible on plain radiographs. In an additional patient, we found that a stress-relieving, pilot holes-only technique did not prevent VB-SFs. The 5 patients operated on with the standard technique composed the background series against which we compared the incidence of VB-SFs in patients operated on with a modification of the standard US technique-a combination of stress-relieving pilot holes, removal of cortex in the chisel path, and a fenestrated chisel (PH/CR/FC)-intended to reduce the incidence of VB-SFs in multilevel PD-L constructs.

Methods: Patients receiving multilevel PD-L implants at 2 sites-1 in the United States and the other in Germany-were operated on with the PH/CR/FC technique and their postoperative CT scans evaluated for the presence of VB-SFs. The frequency of VB-SFs in these patients was compared with that of the 5 patients from the background series who were operated on by the standard US technique. The groups' mean sex, age, body mass index, and vertebral body height, as well as average spinal T score, were also compared.

Results: No fractures were found in 13 interposed vertebral bodies in 11 patients operated on with the PH/CR/FC technique, as compared with 4 VB-SFs and 1 anterior keel cut-to-anterior keel cut fracture in 5 interposed vertebral bodies in 5 patients operated on with the US technique (P ≤ .001). Although the sample sizes were small, this difference in fracture rate was not associated with sex, age, body mass index, or average spinal T score. At up to 13 months of follow-up of patients in the background series, we found that VB-SFs tend not to bridge with bone, instead forming sclerotic margins.

Conclusions: The PH/CR/FC technique studied reduced the incidence of VB-SF in multilevel PD-L implants. Because previously published multilevel studies did not use postoperative CT scans and because VB-SFs are not visible on conventional radiography, the incidence of VB-SFs in multilevel PD-L applications may be higher than previously reported. Our findings may contribute to prevention of complications in total disc replacement.

Citing Articles

ICR in human cadaveric specimens: An essential parameter to consider in a new lumbar disc prosthesis design.

Vanaclocha-Saiz A, Atienza C, Vanaclocha V, Belloch V, Santabarbara J, Jorda-Gomez P N Am Spine Soc J. 2022; 2:100016.

PMID: 35141586 PMC: 8820058. DOI: 10.1016/j.xnsj.2020.100016.


We Need to Talk about Lumbar Total Disc Replacement.

Beatty S Int J Spine Surg. 2018; 12(2):201-240.

PMID: 30276080 PMC: 6159637. DOI: 10.14444/5029.


Physical capability outcomes after total disc replacement with ProDisc-L.

Sullivan H, Bobenmoyer R, Boland K, Cerniglia M, McHugh V, Born H Int J Spine Surg. 2015; 6:43-8.

PMID: 25694870 PMC: 4300876. DOI: 10.1016/j.ijsp.2011.11.001.

References
1.
Hannibal M, Thomas D, Low J, Hsu K, Zucherman J . ProDisc-L total disc replacement: a comparison of 1-level versus 2-level arthroplasty patients with a minimum 2-year follow-up. Spine (Phila Pa 1976). 2007; 32(21):2322-6. DOI: 10.1097/BRS.0b013e3181557c06. View

2.
Bertagnoli R, Yue J, Kershaw T, Shah R, Pfeiffer F, Fenk-Mayer A . Lumbar total disc arthroplasty utilizing the ProDisc prosthesis in smokers versus nonsmokers: a prospective study with 2-year minimum follow-up. Spine (Phila Pa 1976). 2006; 31(9):992-7. DOI: 10.1097/01.brs.0000214970.07626.68. View

3.
Zigler J, Delamarter R, Spivak J, Linovitz R, Danielson 3rd G, Haider T . Results of the prospective, randomized, multicenter Food and Drug Administration investigational device exemption study of the ProDisc-L total disc replacement versus circumferential fusion for the treatment of 1-level degenerative disc disease. Spine (Phila Pa 1976). 2007; 32(11):1155-62. DOI: 10.1097/BRS.0b013e318054e377. View

4.
Shim C, Lee S, Maeng D, Lee S . Vertical split fracture of the vertebral body following total disc replacement using ProDisc: report of two cases. J Spinal Disord Tech. 2005; 18(5):465-9. DOI: 10.1097/01.bsd.0000159035.35365.df. View

5.
Delamarter R, Zigler J, Balderston R, Cammisa F, Goldstein J, Spivak J . Prospective, randomized, multicenter Food and Drug Administration investigational device exemption study of the ProDisc-L total disc replacement compared with circumferential arthrodesis for the treatment of two-level lumbar degenerative disc.... J Bone Joint Surg Am. 2011; 93(8):705-15. DOI: 10.2106/JBJS.I.00680. View