Central and Peripheral Corneal Thickness Measurement in Normal and Keratoconic Eyes Using Three Corneal Pachymeters

Overview
Specialty Ophthalmology
Date 2015 Feb 11
PMID 25667728
Citations 14
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To assess the agreement among ultrasonic pachymetry, the Galilei dual Scheimpflug analyzer, and Orbscan II for central and peripheral (Galilei vs. Orbscan) corneal thickness (CCT and PCT) measurement in normal and keratoconic eyes.

Methods: In this prospective study, CCT and PCT were measured in 88 eyes of 88 refractive surgery candidates and 128 eyes of 69 keratoconic patients with ultrasonic pachymetry, the Galilei, and Orbscan II. The readings by the three instruments were compared using one-way analysis of normal variance. Agreement among the three devices was assessed using Pearson and intraclass correlation coefficients, and Bland-Altman plots. The same analyses were employed to evaluate agreement between Galilei and Orbscan II for PCT measurement.

Results: In the normal group, mean CCT was 551.0±39.4, 566.9±33.5, and 565.5±40.9 μm measured by ultrasonic pachymetry, the Galilei, and Orbscan II, respectively (P<0.001). The corresponding figures in the keratoconus group were 492.0±61.7, 502.0±42.1, and 470.6±56.9 μm, respectively (P<0.001). Mean PCT was 612.5±35.3 and 640.9±38.0 μm in the normal group (P<0.001) and 567.6±35.2 and 595.1±41.4 μm in the keratoconus group (P<0.001) by the Galilei and Orbscan II, respectively. CCT and PCT measurements obtained by different devices were significantly correlated in both groups.

Conclusion: To measure CCT, the Galilei and Orbscan II can be used interchangeably in normal eyes, but not in keratoconic eyes. For PCT, there is a systematic error between measurements obtained by the Galilei and Orbscan II. However, it is possible to change optical pachymeter readings into those obtained by ultrasonic pachymetry using a constant.

Citing Articles

Advancements in Hydrogels for Corneal Healing and Tissue Engineering.

Wu K, Qian S, Faucher A, Tran S Gels. 2024; 10(10).

PMID: 39451315 PMC: 11507397. DOI: 10.3390/gels10100662.


Typical localised element-specific finite element anterior eye model.

Towler J, Consejo A, Zhou D, Romano V, Levis H, Boote C Heliyon. 2023; 9(4):e13944.

PMID: 37101628 PMC: 10123217. DOI: 10.1016/j.heliyon.2023.e13944.


Repeatability of corneal pachymetry and epithelial thickness measurements with spectral-domain optical coherence tomography (SD-OCT) and correlation to ocular surface parameters.

Chatzistergiou V, Tzamalis A, Diafas A, Oustoglou E, Mataftsi A, Tsinopoulos I Int Ophthalmol. 2023; 43(9):3139-3148.

PMID: 37097425 DOI: 10.1007/s10792-023-02713-2.


Prevalence and characteristics of dry eye disease in Parkinson's disease: a systematic review and meta-analysis.

Nagino K, Sung J, Oyama G, Hayano M, Hattori N, Okumura Y Sci Rep. 2022; 12(1):18348.

PMID: 36319814 PMC: 9626467. DOI: 10.1038/s41598-022-22037-y.


The Underlying Relationship between Keratoconus and Down Syndrome.

Akoto T, Li J, Estes A, Karamichos D, Liu Y Int J Mol Sci. 2022; 23(18).

PMID: 36142709 PMC: 9503764. DOI: 10.3390/ijms231810796.


References
1.
Buehl W, Stojanac D, Sacu S, Drexler W, Findl O . Comparison of three methods of measuring corneal thickness and anterior chamber depth. Am J Ophthalmol. 2006; 141(1):7-12. DOI: 10.1016/j.ajo.2005.08.048. View

2.
Gherghel D, Hosking S, Mantry S, Banerjee S, Naroo S, Shah S . Corneal pachymetry in normal and keratoconic eyes: Orbscan II versus ultrasound. J Cataract Refract Surg. 2004; 30(6):1272-7. DOI: 10.1016/j.jcrs.2003.11.049. View

3.
Savini G, Carbonelli M, Barboni P, Hoffer K . Repeatability of automatic measurements performed by a dual Scheimpflug analyzer in unoperated and post-refractive surgery eyes. J Cataract Refract Surg. 2011; 37(2):302-9. DOI: 10.1016/j.jcrs.2010.07.039. View

4.
Doughty M, Zaman M . Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol. 2000; 44(5):367-408. DOI: 10.1016/s0039-6257(00)00110-7. View

5.
Rainer G, Findl O, Petternel V, Kiss B, Drexler W, Skorpik C . Central corneal thickness measurements with partial coherence interferometry, ultrasound, and the Orbscan system. Ophthalmology. 2004; 111(5):875-9. DOI: 10.1016/j.ophtha.2003.09.027. View