Low-power Wireless Electronic Capsule for Long-term Gastrointestinal Monitoring

Overview
Journal J Med Syst
Date 2015 Jan 30
PMID 25631843
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Combining ASIC and multiple microsensors low-power wireless electronic capsule was developed for the long-term monitoring of the entire human gastro-intestinal (GI) tract. To meet the system requirements, several low-power designs were used in the wireless electronic capsule. The capsule measured 11 × 22 mm including batteries (45mAh). The capsule system's lifetime was 233 h, and it could meet the requirements of almost all clinical applications. A wireless electronic capsule, portable data recorder, and workstation comprised the human GI physiological parameters monitoring system. In this paper, this system was used in a clinical trial to compare colon peristaltic pressure between patients with constipation and healthy people.

Citing Articles

Smart capsules for sensing and sampling the gut: status, challenges and prospects.

Rehan M, Al-Bahadly I, Thomas D, Young W, Cheng L, Avci E Gut. 2023; 73(1):186-202.

PMID: 37734912 PMC: 10715516. DOI: 10.1136/gutjnl-2023-329614.


A Flexible Implant for Multi-Day Monitoring of Colon Segment Activity.

Majerus S, Cabal D, Hacohen Y, Hanzlicek B, Smiley A, Wang Y IEEE Trans Biomed Circuits Syst. 2023; 17(5):941-951.

PMID: 37363840 PMC: 10732233. DOI: 10.1109/TBCAS.2023.3289768.


A Software Development Platform for Wearable Medical Applications.

Zhang R, Lin W J Med Syst. 2015; 39(10):309.

PMID: 26276017 DOI: 10.1007/s10916-015-0309-0.

References
1.
MacKay R, Jacobson B . Endoradiosonde. Nature. 1957; 179(4572):1239-40. DOI: 10.1038/1791239a0. View

2.
Duroux P, Emde C, Bauerfeind P, Francis C, Grisel A, Thybaud L . The ion sensitive field effect transistor (ISFET) pH electrode: a new sensor for long term ambulatory pH monitoring. Gut. 1991; 32(3):240-5. PMC: 1378826. DOI: 10.1136/gut.32.3.240. View

3.
Pan G, Yan G, Qiu X, Cui J . Bleeding detection in Wireless Capsule Endoscopy based on Probabilistic Neural Network. J Med Syst. 2010; 35(6):1477-84. DOI: 10.1007/s10916-009-9424-0. View

4.
Hammond P, Ali D, Cumming D . A system-on-chip digital pH meter for use in a wireless diagnostic capsule. IEEE Trans Biomed Eng. 2005; 52(4):687-94. DOI: 10.1109/TBME.2005.844041. View

5.
Pye G, Evans D, Ledingham S, Hardcastle J . Gastrointestinal intraluminal pH in normal subjects and those with colorectal adenoma or carcinoma. Gut. 1990; 31(12):1355-7. PMC: 1378756. DOI: 10.1136/gut.31.12.1355. View