Laboratory Blood Analysis in Strigiformes-Part II: Plasma Biochemistry Reference Intervals and Agreement Between the Abaxis Vetscan V2 and the Roche Cobas C501

Overview
Journal Vet Clin Pathol
Date 2015 Jan 24
PMID 25613649
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Limited plasma biochemical information is available in Strigiformes. Only one study investigated the agreement between a point-of-care with a reference laboratory analyzer for biochemistry variables in birds.

Objectives: The objective was to report reference intervals (RI) for plasma biochemistry variables in Strigiformes, and to assess agreement between the Abaxis Vetscan V2 and Roche Cobas c501.

Methods: A prospective study was designed to assess plasma biochemistry RI for concentration of calcium, phosphorus, total protein, albumin, globulin, glucose, bilirubin, uric acid, bile acids, sodium, potassium, and chloride, and activities of AST, GGT, CK, amylase, lipase, LDH, and GLDH. In addition, the agreement between the Vetscan and the Cobas in owl species was assessed.

Results: A total of 190 individuals were sampled belonging to 12 Strigiformes species including Barn Owls, Barred Owls, Great Horned Owls, Eurasian Eagle Owls, Spectacled Owls, Eastern Screech Owls, Long-Eared Owls, Short-Eared Owls, Great Gray Owls, Snowy Owls, Northern Saw-Whet Owls, and Northern Hawk-Owls. Order-, species-, and method-specific RI were determined on both analyzers. Although Vetscan data were not equivalent to the Cobas, 4 analytes (glucose, AST, CK, and total protein, with correction for bias) were within acceptable agreement, 3 analytes (uric acid, calcium, and phosphorus) were within close agreement, and the remaining analytes were in strong disagreement. Species-specific differences were observed notably for the concentration of glucose in Barn Owls and electrolytes in Northern Saw-Whet Owls.

Conclusions: Overall, this study suggests that the Vetscan has acceptable clinical performance in Strigiformes for some analytes and highlights discrepancies for several analytes.

Citing Articles

Reference Intervals for Selected Hematology and Clinical Chemistry Measurands in Temminck's Pangolin ().

Hooijberg E, Lourens K, Meyer L Front Vet Sci. 2021; 8:654529.

PMID: 34307517 PMC: 8295727. DOI: 10.3389/fvets.2021.654529.


Comparison of portable and conventional laboratory analyzers for biochemical tests in chickens.

Ruiz-Jimenez F, Gruber E, Correa M, Crespo R Poult Sci. 2021; 100(2):746-754.

PMID: 33518128 PMC: 7858187. DOI: 10.1016/j.psj.2020.11.060.


A fatal case of a captive snowy owl (Bubo scandiacus) with Haemoproteus infection in Japan.

Yoshimoto M, Ozawa K, Kondo H, Echigoya Y, Shibuya H, Sato Y Parasitol Res. 2020; 120(1):277-288.

PMID: 33191448 DOI: 10.1007/s00436-020-06972-1.


Biochemical reference intervals for homing pigeons in China.

Zheng M, Lv Q, Fu X, Shi F Poult Sci. 2020; 99(7):3463-3468.

PMID: 32616240 PMC: 7597933. DOI: 10.1016/j.psj.2020.04.022.


Hematologic and serum biochemical reference intervals of the Oriental white stork (Ciconia boyciana) and the application of an automatic hematologic analyzer.

Han J, Jang H, Na K J Vet Sci. 2016; 17(3):399-405.

PMID: 26726025 PMC: 5037309. DOI: 10.4142/jvs.2016.17.3.399.