» Articles » PMID: 25502644

ISCEV Standard for Full-field Clinical Electroretinography (2015 Update)

Overview
Journal Doc Ophthalmol
Specialty Ophthalmology
Date 2014 Dec 16
PMID 25502644
Citations 684
Authors
Affiliations
Soon will be listed here.
Abstract

This document, from the International Society for Clinical Electrophysiology of Vision (ISCEV), presents an updated and revised ISCEV Standard for full-field clinical electroretinography (ffERG or simply ERG). The parameters for Standard flash stimuli have been revised to accommodate a variety of light sources including gas discharge lamps and light emitting diodes. This ISCEV Standard for clinical ERGs specifies six responses based on the adaptation state of the eye and the flash strength: (1) Dark-adapted 0.01 ERG (rod ERG); (2) Dark-adapted 3 ERG (combined rod-cone standard flash ERG); (3) Dark-adapted 3 oscillatory potentials; (4) Dark-adapted 10 ERG (strong flash ERG); (5) Light-adapted 3 ERG (standard flash "cone" ERG); and (6) Light-adapted 30 Hz flicker ERG. ISCEV encourages the use of additional ERG protocols for testing beyond this minimum standard for clinical ERGs.

Citing Articles

SLC37A3-associated retinitis pigmentosa: a case report of clinical features and three-year follow up.

Kong M, Bailey J, Exinor A, Piamjitchol C, Demirkol A, Tsang S Doc Ophthalmol. 2025; .

PMID: 40069519 DOI: 10.1007/s10633-025-10013-6.


Comparison of ERG signal-to-noise ratios in simultaneous recordings with skin electrodes and contact lens electrodes.

Brodie S, Faruque P, Pincay J, Sylla M, Cui X, Choi S Doc Ophthalmol. 2025; .

PMID: 39903329 DOI: 10.1007/s10633-025-10003-8.


Atypical Leber Hereditary Optic Neuropathy (LHON) Associated with a Novel MT-CYB:m.15309T>C(Ile188Thr) Variant.

Petrovic Pajic S, Fakin A, Jarc-Vidmar M, Sustar Habjan M, Malinar L, Pavlovic K Genes (Basel). 2025; 16(1).

PMID: 39858655 PMC: 11764998. DOI: 10.3390/genes16010108.


Longitudinal assessment of retinal and visual pathway electrophysiology and structure after high altitude exposure.

Shi X, Li M, Zhang X, Yuan F, Liu Y, Lin J Graefes Arch Clin Exp Ophthalmol. 2025; .

PMID: 39820514 DOI: 10.1007/s00417-024-06729-x.


Rescue of Aberrant Splicing Caused by a Novel Complex Deep-intronic Allele.

Maggi J, Feil S, Gloggnitzer J, Maggi K, Hanson J, Koller S Genes (Basel). 2025; 15(12.

PMID: 39766771 PMC: 11675205. DOI: 10.3390/genes15121503.


References
1.
Hood D, Bach M, Brigell M, Keating D, Kondo M, Lyons J . ISCEV standard for clinical multifocal electroretinography (mfERG) (2011 edition). Doc Ophthalmol. 2011; 124(1):1-13. PMC: 4466109. DOI: 10.1007/s10633-011-9296-8. View

2.
Marmor M, Brigell M, McCulloch D, Westall C, Bach M . ISCEV standard for clinical electro-oculography (2010 update). Doc Ophthalmol. 2011; 122(1):1-7. DOI: 10.1007/s10633-011-9259-0. View

3.
Marmor M, Fulton A, Holder G, Miyake Y, Brigell M, Bach M . ISCEV Standard for full-field clinical electroretinography (2008 update). Doc Ophthalmol. 2008; 118(1):69-77. DOI: 10.1007/s10633-008-9155-4. View

4.
Odom J, Bach M, Brigell M, Holder G, McCulloch D, Tormene A . ISCEV standard for clinical visual evoked potentials (2009 update). Doc Ophthalmol. 2009; 120(1):111-9. DOI: 10.1007/s10633-009-9195-4. View

5.
Brigell M, Bach M, Barber C, Moskowitz A, Robson J . Guidelines for calibration of stimulus and recording parameters used in clinical electrophysiology of vision. Doc Ophthalmol. 2003; 107(2):185-93. DOI: 10.1023/a:1026244901657. View