Monte Carlo Simulation of a 6 MV X-Ray Beam for Open and Wedge Radiation Fields, Using GATE Code

Overview
Date 2014 Nov 27
PMID 25426430
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

The aim of this study is to provide a control software system, based on Monte Carlo simulation, and calculations of dosimetric parameters of standard and wedge radiation fields, using a Monte Carlo method. GATE version 6.1 (OpenGATE Collaboration), was used to simulate a compact 6 MV linear accelerator system. In order to accelerate the calculations, the phase-space technique and cluster computing (Condor version 7.2.4, Condor Team, University of Wisconsin-Madison) were used. Dosimetric parameters used in treatment planning systems for the standard and wedge radiation fields (10 cm × 10 cm to 30 cm × 30 cm and a 60° wedge), including the percentage depth dose and dose profiles, were measured by both computational and experimental methods. Gamma index was applied to compare calculated and measured results with 3%/3 mm criteria. Gamma index was applied to compare calculated and measured results. Almost all calculated data points have satisfied gamma index criteria of 3% to 3 mm. Based on the good agreement between calculated and measured results obtained for various radiation fields in this study, GATE may be used as a useful tool for quality control or pretreatment verification procedures in radiotherapy.

Citing Articles

Determination and validation of the initial beam parameters of Elekta Agility collimator head by Monte Carlo simulations.

Chand B, Singh R, Kumar M Phys Eng Sci Med. 2022; 45(3):889-899.

PMID: 35849322 DOI: 10.1007/s13246-022-01159-7.

References
1.
Schmidtlein C, Kirov A, Nehmeh S, Erdi Y, Humm J, Amols H . Validation of GATE Monte Carlo simulations of the GE Advance/Discovery LS PET scanners. Med Phys. 2006; 33(1):198-208. DOI: 10.1118/1.2089447. View

2.
Thiam C, Breton V, Donnarieix D, Habib B, Maigne L . Validation of a dose deposited by low-energy photons using GATE/GEANT4. Phys Med Biol. 2008; 53(11):3039-55. DOI: 10.1088/0031-9155/53/11/019. View

3.
Jabbari K, Roayaei M, Saberi H . Calculation of excess dose to the eye phantom due to a distanced shielding for electron therapy in head and neck cancers. J Med Signals Sens. 2013; 2(3):144-8. PMC: 3660709. View

4.
Grevillot L, Bertrand D, Dessy F, Freud N, Sarrut D . A Monte Carlo pencil beam scanning model for proton treatment plan simulation using GATE/GEANT4. Phys Med Biol. 2011; 56(16):5203-19. DOI: 10.1088/0031-9155/56/16/008. View

5.
Benhalouche S, Visvikis D, Le Maitre A, Pradier O, Boussion N . Evaluation of clinical IMRT treatment planning using the GATE Monte Carlo simulation platform for absolute and relative dose calculations. Med Phys. 2013; 40(2):021711. DOI: 10.1118/1.4774358. View