Purpose: To develop a perimetric test strategy, Structure Estimation of Minimum Uncertainty (SEMU), that uses structural information to drive stimulus choices.

Methods: Structure Estimation of Minimum Uncertainty uses retinal nerve fiber layer (RNFL) thickness data as measured by optical coherence tomography to predict perimetric sensitivity. This prediction is used to set suprathreshold levels that then alter a prior probability distribution of the final test output. Using computer simulation, we studied SEMU's performance under three different patient error response conditions: No Error, Typical False Positive errors, and Extremely Unreliable patients. In experiment 1, SEMU was compared with an existing suprathreshold cum thresholding combination test procedure, Estimation of Minimum Uncertainty (EMU), on single visual field locations. We used these results to finalize SEMU parameters. In experiment 2, SEMU was compared with full threshold (FT) on 163 glaucomatous visual fields.

Results: On individual locations, SEMU has similar accuracy to EMU, but is, on average, one presentation faster than EMU. For the typical false-positive error condition, SEMU has significantly lower error compared with FT (SEMU average 0.33 dB lower; p < 0.001) and the 90% measured sensitivity range for SEMU is also smaller than that for FT. For unreliable patients, however, FT has lower mean and SD of error. Structure Estimation of Minimum Uncertainty makes significantly fewer presentations than FT (1.08 presentation on average fewer in a typical false-positive condition; p < 0.001). Assuming that a location in the field is marked abnormal if it falls below the 5th percentile of normal, SEMU has a false-positive rate of less than 10% for all error conditions compared with FT's rate of 20% or more.

Conclusions: On average, simulations show that using RNFL information to guide stimulus placement in a perimetric test procedure maintains accuracy, improves precision, and decreases test duration for patients with less than 15% false-positive rates.

Citing Articles

Frequency-of-seeing curves (psychometric functions) for perimetric stimuli in age-related macular degeneration.

Denniss J, Baggaley H, Astle A Ophthalmic Physiol Opt. 2024; 45(1):301-307.

PMID: 39329222 PMC: 11629837. DOI: 10.1111/opo.13396.


A Practical Framework for the Integration of Structural Data Into Perimetric Examinations.

Evans J, Ometto G, Crabb D, Montesano G Transl Vis Sci Technol. 2024; 13(6):19.

PMID: 38916881 PMC: 11205229. DOI: 10.1167/tvst.13.6.19.


Improving the Accuracy and Speed of Visual Field Testing in Glaucoma With Structural Information and Deep Learning.

Montesano G, Lazaridis G, Ometto G, Crabb D, Garway-Heath D Transl Vis Sci Technol. 2023; 12(10):10.

PMID: 37831447 PMC: 10587851. DOI: 10.1167/tvst.12.10.10.


Suprathreshold Approaches to Mapping the Visual Field in Advanced Glaucoma.

Denniss J, McKendrick A, Turpin A Transl Vis Sci Technol. 2023; 12(6):19.

PMID: 37358492 PMC: 10297785. DOI: 10.1167/tvst.12.6.19.


The Usefulness of Assessing Glaucoma Progression With Postprocessed Visual Field Data.

Abu S, Poleon S, Racette L Transl Vis Sci Technol. 2022; 11(5):5.

PMID: 35522306 PMC: 9078054. DOI: 10.1167/tvst.11.5.5.