A Novel Method to Assess Primary Stability of Press-fit Acetabular Cups

Overview
Date 2014 Nov 12
PMID 25384445
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Initial stability is an essential prerequisite to achieve osseointegration of press-fit acetabular cups in total hip replacements. Most in vitro methods that assess cup stability do not reproduce physiological loading conditions and use simplified acetabular models with a spherical cavity. The aim of this study was to investigate the effect of bone density and acetabular geometry on cup stability using a novel method for measuring acetabular cup micromotion. A press-fit cup was inserted into Sawbones(®) foam blocks having different densities to simulate normal and osteoporotic bone variations and different acetabular geometries. The stability of the cup was assessed in two ways: (a) measurement of micromotion of the cup in 6 degrees of freedom under physiological loading and (b) uniaxial push-out tests. The results indicate that changes in bone substrate density and acetabular geometry affect the stability of press-fit acetabular cups. They also suggest that cups implanted into weaker, for example, osteoporotic, bone are subjected to higher levels of micromotion and are therefore more prone to loosening. The decrease in stability of the cup in the physiological model suggests that using simplified spherical cavities to model the acetabulum over-estimates the initial stability of press-fit cups. This novel testing method should provide the basis for a more representative protocol for future pre-clinical evaluation of new acetabular cup designs.

Citing Articles

Dynamics of manual impaction instruments during total hip arthroplasty.

Reynolds A, Doyle R, Boughton O, Cobb J, Muirhead-Allwood S, Jeffers J Bone Joint Res. 2024; 13(4):193-200.

PMID: 38649151 PMC: 11041854. DOI: 10.1302/2046-3758.134.BJR-2023-0224.R1.


A novel primary stability test method for artificial acetabular shells considering vertical load during level walking and shell position.

Yoshida K, Fukushima K, Sakai R, Uchiyama K, Takahira N, Ujihira M PLoS One. 2024; 19(2):e0296919.

PMID: 38421998 PMC: 10903903. DOI: 10.1371/journal.pone.0296919.


Implications of humeral short-stem diametral sizing on implant stability.

Reeves J, Spangenberg G, Elwell J, Stewart B, Vanasse T, Roche C JSES Int. 2023; 7(6):2445-2453.

PMID: 37969503 PMC: 10638575. DOI: 10.1016/j.jseint.2023.06.023.


Comparison of Test Setups for the Experimental Evaluation of the Primary Fixation Stability of Acetabular Cups.

Schulze C, Vogel D, Mallow S, Bader R Materials (Basel). 2020; 13(18).

PMID: 32916802 PMC: 7559462. DOI: 10.3390/ma13183982.


Impaction technique influences implant stability in low-density bone model.

Doyle R, Van Arkel R, Muirhead-Allwood S, Jeffers J Bone Joint Res. 2020; 9(7):386-393.

PMID: 32793333 PMC: 7393184. DOI: 10.1302/2046-3758.97.BJR-2019-0303.R1.