» Articles » PMID: 25258499

Comparison of the Cheese-wiring Effects Among Three Sutures Used in Rotator Cuff Repair

Overview
Date 2014 Sep 27
PMID 25258499
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The goal of this study was to compare the cheese-wiring effects of three sutures with different coefficients of friction.

Materials And Methods: Sixteen human cadaveric shoulders were dissected to expose the distal supraspinatus and infraspinatus muscle tendons. Three sutures were stitched through the tendons: #2 Orthocord(™) suture (reference #223114, DePuy Mitek, Inc., Raynham, MA), #2 ETHIBOND* EXCEL Suture, and #2 FiberWire(®) suture (FiberWire(®), Arthrex, Naples, FL). The sutures were pulled by cyclic axial forces from 10 to 70 N at 1 Hz for 1000 cycles through a MTS machine. The cut-through distance on the tendon was measured with a digital caliper.

Results: The cut-through distance in the supraspinatus tendons (mean ± standard deviation, n = 12) were 2.9 ± 0.6 mm for #2 Orthocord(™) suture, 3.2 ± 1.2 mm for #2 ETHIBOND* suture, and 4.2 ± 1.7 mm for #2 FiberWire(®) suture. The differences were statistically significant analyzing with analysis of variance (P = 0.047) and two-tailed Student's t-test, which showed significance between Orthocord(™) and FiberWire(®) sutures (P = 0.026), but not significant between Orthocord(™) and ETHIBOND* sutures (P = 0.607) or between ETHIBOND* and FiberWire(®) sutures (P = 0.103).

Conclusion: The cheese-wiring effect is less in the Orthocord(™) suture than in the FiberWire(®) suture in human cadaveric supraspinatus tendons.

Clinical Relevance: Identification of sutures that cause high levels of tendon cheese-wiring after rotator cuff repair can lead to better suture selection.

Citing Articles

Whipstitch and Locking Stitch Show Equivalent Elongation and Load to Failure Across 3 Suture Systems in a Biomechanical Model of Quadriceps Tendon Grafts for Anterior Cruciate Ligament Reconstruction.

Diaz M, Branch E, Dunn J, Brothers A, Jordan S Arthrosc Sports Med Rehabil. 2024; 6(5):100968.

PMID: 39534030 PMC: 11551357. DOI: 10.1016/j.asmr.2024.100968.


A Dynamic, Self-Tensioning Suture Contracts in Saline to Counteract Changes in Loop Length From Cyclic Loading.

Mehta N, Shewman E, Allahabadi S, Liu S, Lawande N, Dunlap B Arthrosc Sports Med Rehabil. 2024; 6(4):100872.

PMID: 39421349 PMC: 11480785. DOI: 10.1016/j.asmr.2023.100872.


A highly elastic absorbable monofilament suture fabricated from poly(3-hydroxybutyrate-co-4-hydroxybutyrate).

Murayama A, Yoneda H, Maehara A, Shiomi N, Hirata H Sci Rep. 2023; 13(1):3275.

PMID: 36841914 PMC: 9968320. DOI: 10.1038/s41598-023-30292-w.


A Weaving Rip-Stop Technique Leads to a Significantly Increased Load to Failure and Reduction in Suture-Tendon Cut-Through in a Biomechanical Model of Rotator Cuff Repair.

Neeley R, Diaz M, Gorman 2nd R, Frankle M, Mighell M Arthrosc Sports Med Rehabil. 2021; 3(5):e1263-e1272.

PMID: 34712962 PMC: 8527263. DOI: 10.1016/j.asmr.2021.04.005.


Rotator cuff retear after repair surgery: comparison between experienced and inexperienced surgeons.

Park J, Lee J, Oh K, Chung S, Choi Y, Yoon W Clin Shoulder Elb. 2021; 24(3):135-140.

PMID: 34488293 PMC: 8423529. DOI: 10.5397/cise.2021.00073.


References
1.
Wust D, Meyer D, Favre P, Gerber C . Mechanical and handling properties of braided polyblend polyethylene sutures in comparison to braided polyester and monofilament polydioxanone sutures. Arthroscopy. 2006; 22(11):1146-53. DOI: 10.1016/j.arthro.2006.06.013. View

2.
Wright P, Budoff J, Yeh M, Kelm Z, Luo Z . Strength of damaged suture: an in vitro study. Arthroscopy. 2006; 22(12):1270-1275.e3. DOI: 10.1016/j.arthro.2006.08.019. View

3.
Swan Jr K, Baldini T, McCarty E . Arthroscopic suture material and knot type: an updated biomechanical analysis. Am J Sports Med. 2009; 37(8):1578-85. DOI: 10.1177/0363546509332816. View

4.
Williams Jr G, Rockwood Jr C, Bigliani L, Iannotti J, Stanwood W . Rotator cuff tears: why do we repair them?. J Bone Joint Surg Am. 2004; 86(12):2764-76. View

5.
Mahar A, Moezzi D, Serra-Hsu F, Pedowitz R . Comparison and performance characteristics of 3 different knots when tied with 2 suture materials used for shoulder arthroscopy. Arthroscopy. 2006; 22(6):614.e1-2. DOI: 10.1016/j.arthro.2006.02.005. View