Is Peracetic Acid Suitable for the Cleaning Step of Reprocessing Flexible Endoscopes?

Overview
Date 2014 Sep 18
PMID 25228941
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

The bioburden (blood, protein, pathogens and biofilm) on flexible endoscopes after use is often high and its removal is essential to allow effective disinfection, especially in the case of peracetic acid-based disinfectants, which are easily inactivated by organic material. Cleaning processes using conventional cleaners remove a variable but often sufficient amount of the bioburden. Some formulations based on peracetic acid are recommended by manufacturers for the cleaning step. We performed a systematic literature search and reviewed the available evidence to clarify the suitability of peracetic acid-based formulations for cleaning flexible endoscopes. A total of 243 studies were evaluated. No studies have yet demonstrated that peracetic acid-based cleaners are as effective as conventional cleaners. Some peracetic acid-based formulations have demonstrated some biofilm-cleaning effects and no biofilm-fixation potential, while others have a limited cleaning effect and a clear biofilm-fixation potential. All published data demonstrated a limited blood cleaning effect and a substantial blood and nerve tissue fixation potential of peracetic acid. No evidence-based guidelines on reprocessing flexible endoscopes currently recommend using cleaners containing peracetic acid, but some guidelines clearly recommend not using them because of their fixation potential. Evidence from some outbreaks, especially those involving highly multidrug-resistant gram-negative pathogens, indicated that disinfection using peracetic acid may be insufficient if the preceding cleaning step is not performed adequately. Based on this review we conclude that peracetic acid-based formulations should not be used for cleaning flexible endoscopes.

Citing Articles

.

Bundesgesundheitsblatt Gesundheitsforschung Gesundheitsschutz. 2024; 67(12):1410-1468.

PMID: 39352496 DOI: 10.1007/s00103-024-03942-1.


Quantification of peracetic acid (PAA) in the HO + acetic acid reaction by the wavelength shift analysis in near-UV/visible absorption region.

Chen G, Lin Y, Fu C, Lin C, Muthiah B, Espulgar W Anal Sci. 2024; 40(3):489-499.

PMID: 38165524 DOI: 10.1007/s44211-023-00481-8.


Susceptibility of Spores to Selected Reference Substances and Disinfectants.

Chojecka A Pol J Microbiol. 2022; 71(3):353-358.

PMID: 36185021 PMC: 9608164. DOI: 10.33073/pjm-2022-031.


Does the Reprocessing of Endoscopes Have to Take Place Immediately after Pre-Cleaning? A First Evaluation.

Eichel V, Jabs J, Unser S, Mutters N, Scherrer M Clin Endosc. 2021; 54(4):526-533.

PMID: 34256556 PMC: 8357578. DOI: 10.5946/ce.2020.238.


Virucidal efficacy of peracetic acid for instrument disinfection.

Becker B, Brill F, Todt D, Steinmann E, Lenz J, Paulmann D Antimicrob Resist Infect Control. 2017; 6:114.

PMID: 29177047 PMC: 5693215. DOI: 10.1186/s13756-017-0271-3.


References
1.
Balsamo A, Uchikawa Graziano K, Schneider R, Antunes Junior M, Lacerda R . [Removing biofilm from a endoscopic: evaluation of disinfection methods currently used]. Rev Esc Enferm USP. 2012; 46 Spec No:91-8. DOI: 10.1590/s0080-62342012000700014. View

2.
Hanson P, Gor D, Jeffries D, Collins J . Elimination of high titre HIV from fibreoptic endoscopes. Gut. 1990; 31(6):657-9. PMC: 1378490. DOI: 10.1136/gut.31.6.657. View

3.
Spaun G, Goers T, Pierce R, Cassera M, Scovil S, Swanstrom L . Use of flexible endoscopes for NOTES: sterilization or high-level disinfection?. Surg Endosc. 2009; 24(7):1581-8. DOI: 10.1007/s00464-009-0815-6. View

4.
Hernandez A, Martro E, Puzo C, Matas L, Burgues C, Vazquez N . In-use evaluation of Perasafe compared with Cidex in fibreoptic bronchoscope disinfection. J Hosp Infect. 2003; 54(1):46-51. DOI: 10.1016/s0195-6701(03)00072-0. View

5.
Burdick J, Hambrick D . Endoscope reprocessing and repair costs. Gastrointest Endosc Clin N Am. 2004; 14(4):717-24, ix-x. DOI: 10.1016/j.giec.2004.05.002. View