Clinical Validation of a Low-power and Wearable ECG Patch for Long Term Full-disclosure Monitoring

Overview
Date 2014 Sep 10
PMID 25201418
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Detection of intermittent atrial fibrillation (AF) is done using a 24-h Holter. Holter recordings are powerful but lack the comfort and have limited recording times resulting in under diagnosing of intermittent AF.

Objective: Within this work we evaluated and compared a novel miniaturized three-channel ECG monitoring patch versus a 24-h Holter system.

Methods: Both patients with a chronic AF rhythm (n=5) as well as patients with an AF rhythm that underwent electrical reconversion (n = 5) were equipped with both a 24-h Holter and ECG patch.

Results: Alignment of raw data of both ECG systems allowed cross-correlation analysis. Overall good correlations of up to 85% were obtained. RR-interval analysis of both systems resulted in very high correlations of 99% and higher. AF analysis showed correct identification of AF on both ECG systems.

Conclusions: The performance of our ECG patch matches that of the 24-h Holter and could provide a suitable tool for long-term monitoring applications.

Citing Articles

Reliability and validity of a novel single-lead portable electrocardiogram device for pregnant women: a comparative study.

Wang H, Wang J, Jing W, Dai S, Zhang D, Geng S BMC Med Inform Decis Mak. 2025; 25(1):108.

PMID: 40033253 PMC: 11877860. DOI: 10.1186/s12911-025-02952-6.


Detection of Clenbuterol-Induced Changes in Heart Rate Using At-Home Recorded Smartwatch Data: Randomized Controlled Trial.

Elzinga W, Prins S, Borghans L, Gal P, Vargas G, Groeneveld G JMIR Form Res. 2021; 5(12):e31890.

PMID: 34967757 PMC: 8759015. DOI: 10.2196/31890.


Cardiac Ambulatory Monitoring: New Wireless Device Validated Against Conventional Holter Monitoring in a Case Series.

Murali S, Brugger N, Rincon F, Mashru M, Cook S, Goy J Front Cardiovasc Med. 2020; 7:587945.

PMID: 33330650 PMC: 7733961. DOI: 10.3389/fcvm.2020.587945.


Reconstruction of 12-Lead Electrocardiogram from a Three-Lead Patch-Type Device Using a LSTM Network.

Sohn J, Yang S, Lee J, Ku Y, Kim H Sensors (Basel). 2020; 20(11).

PMID: 32526828 PMC: 7309162. DOI: 10.3390/s20113278.


Windows Into Human Health Through Wearables Data Analytics.

Witt D, Kellogg R, Snyder M, Dunn J Curr Opin Biomed Eng. 2019; 9:28-46.

PMID: 31832566 PMC: 6907085. DOI: 10.1016/j.cobme.2019.01.001.