Influence of Skin Colour on Diagnostic Accuracy of the Jaundice Meter JM 103 in Newborns

Overview
Date 2014 Jul 31
PMID 25074981
Citations 13
Authors
Affiliations
Soon will be listed here.
Abstract

Aim: To assess the diagnostic accuracy of the JM 103 as a screening tool for neonatal jaundice and explore differential effects based on skin colour.

Methods: We prospectively compared the transcutaneous bilirubin (TcB) and serum bilirubin (TSB) measurements of newborns over a 3 month-period. Skin colour was assigned via reference colour swatches. Diagnostic measures of the TcB/TSB comparison were made and clinically relevant TcB cut-off values were determined for each skin colour group.

Results: 451 infants (51 light, 326 medium and 74 dark skin colour) were recruited. The association between TcB and TSB was high for all skin colours (rs>0.9). The Bland-Altman analysis showed an absolute mean difference between the two measures of 13.3±26.4 µmol/L with broad limits of agreement (-39.4-66.0 µmol/L), with TcB underestimating TSB in light and medium skin colours and overestimating in dark skin colour. Diagnostic measures were also consistently high across skin colours, with no clinically significant differences observed.

Conclusions: The JM 103 is a useful screening tool to identify infants in need of serum bilirubin, regardless of skin colour. The effect of skin colour on the accuracy of this device at high levels of serum bilirubin could not be assessed fully due to small numbers in the light and dark groups.

Citing Articles

Transcutaneous bilirubin reliability during and after phototherapy depending on skin color.

Candel-Pau J, Maya-Enero S, Garcia-Garcia J, Duran-Jorda X, Lopez-Vilchez M Eur J Pediatr. 2024; 183(7):2819-2830.

PMID: 38581462 PMC: 11192662. DOI: 10.1007/s00431-024-05516-4.


Transcutaneous bilirubinometry versus total serum bilirubin measurement for newborns.

Okwundu C, Olowoyeye A, Uthman O, Smith J, Wiysonge C, Bhutani V Cochrane Database Syst Rev. 2023; 5:CD012660.

PMID: 37158489 PMC: 10167941. DOI: 10.1002/14651858.CD012660.pub2.


Handheld diffuse reflectance spectroscopy system for noninvasive quantification of neonatal bilirubin and hemoglobin concentrations: a pilot study.

Cheng N, Tzeng S, Fang M, Kuo C, Lu W, Yang C Biomed Opt Express. 2023; 14(1):467-476.

PMID: 36698673 PMC: 9842000. DOI: 10.1364/BOE.475531.


Quantification of cephalocaudal progression of jaundice in preterm infants.

Dam-Vervloet A, de Boer F, Nijholt I, Poot L, Bosschaart N, van Straaten H Pediatr Res. 2022; 94(1):239-245.

PMID: 36443401 PMC: 10356589. DOI: 10.1038/s41390-022-02396-y.


The Impact of Phototherapy on the Accuracy of Transcutaneous Bilirubin Measurements in Neonates: Optimal Measurement Site and Timing.

Ho S, Lin Y, Chen C Diagnostics (Basel). 2021; 11(9).

PMID: 34574069 PMC: 8466921. DOI: 10.3390/diagnostics11091729.