Evaluation of the Bruker Biotyper and VITEK MS MALDI-TOF MS Systems for the Identification of Unusual And/or Difficult-to-identify Microorganisms Isolated from Clinical Specimens

Overview
Publisher Springer
Date 2014 Jun 26
PMID 24962194
Citations 37
Authors
Affiliations
Soon will be listed here.
Abstract

The purpose of this investigation was to evaluate the analytical performance characteristics of matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) for the identification of unusual organisms. We evaluated the accuracy of two MALDI-TOF MS systems, bioMérieux VITEK MS (database v2.0) and Bruker Biotyper (software version 3.0), for the identification of the most difficult and/or unusual microorganisms isolated from clinical specimens. Our study included 174 bacterial isolates recovered from clinical cultures at Barnes-Jewish Hospital, St. Louis, MO, from 2009 to 2013, representing 50 genera and 52 species. MS identifications were compared to the identification reported by the reference laboratory. Discrepancies were resolved using molecular methods, including 16S rRNA gene sequencing and additional molecular methods. When performed, molecular methods were considered the gold standard. Of the 168 isolates resolved to the genus level, VITEK MS identified 145 (86.3 %), and of the 114 isolates resolved to the species level, 97 (85.1 %) were correctly identified. Bruker Biotyper identified 155 (92.3 %) of 168 isolates to the genus level and 97 (85.1 %) of 114 isolates to the species level. VITEK MS and Bruker Biotyper provided no identification for 17 (10.1 %) and 12 (7.1 %) organisms, respectively, and misidentified six (3.6 %) and one (0.6 %) isolate, respectively. Six isolates (3.6 %) were not resolvable to the genus level and were excluded from data analysis due to the lack of a gold standard for comparison. There was no significant difference in the number of organisms identified to the genus level, species level, unidentified, or misidentified by the two MALDI-TOF MS systems (p = 0.11, 1.0, 0.44, and 0.12, respectively).

Citing Articles

Analysis of the efficacy of MALDI-TOF MS technology in identifying microorganisms in cancer patients and oncology hospital environment.

Czeszewska-Rosiak G, Adamczyk I, Ludwiczak A, Fijalkowski P, Fijalkowski P, Twaruzek M Heliyon. 2025; 11(2):e42015.

PMID: 39906802 PMC: 11791110. DOI: 10.1016/j.heliyon.2025.e42015.


Enhanced vitamin B production by isolated Bacillus strains with the application of response surface methodology.

Abdel-Baki R, Ahmed M, Barakat O, Khalafalla G BMC Biotechnol. 2024; 24(1):90.

PMID: 39529053 PMC: 11555979. DOI: 10.1186/s12896-024-00919-5.


Myroides species, pathogenic spectrum and clinical microbiology sight in Mexican isolates.

Colin-Castro C, Ortiz-Alvarez J, Hernandez-Perez C, Hernandez-Duran M, Garcia-Hernandez M, Martinez-Zavaleta M PLoS One. 2024; 19(11):e0310262.

PMID: 39495724 PMC: 11534234. DOI: 10.1371/journal.pone.0310262.


A narrative review of wastewater surveillance: pathogens of concern, applications, detection methods, and challenges.

Singh S, Ismail Ahmed A, Almansoori S, Alameri S, Adlan A, Odivilas G Front Public Health. 2024; 12:1445961.

PMID: 39139672 PMC: 11319304. DOI: 10.3389/fpubh.2024.1445961.


Characterization by MALDI-TOF MS and 16S rRNA Gene Sequencing of Aerobic Endospore-Forming Bacteria Isolated from Pharmaceutical Facility in Rio de Janeiro, Brazil.

Caldeira N, de Souza M, Miranda R, Costa L, Forsythe S, Zahner V Microorganisms. 2024; 12(4).

PMID: 38674668 PMC: 11051829. DOI: 10.3390/microorganisms12040724.


References
1.
Cherkaoui A, Hibbs J, Emonet S, Tangomo M, Girard M, Francois P . Comparison of two matrix-assisted laser desorption ionization-time of flight mass spectrometry methods with conventional phenotypic identification for routine identification of bacteria to the species level. J Clin Microbiol. 2010; 48(4):1169-75. PMC: 2849558. DOI: 10.1128/JCM.01881-09. View

2.
Fang H, Ohlsson A, Ullberg M, Ozenci V . Evaluation of species-specific PCR, Bruker MS, VITEK MS and the VITEK 2 system for the identification of clinical Enterococcus isolates. Eur J Clin Microbiol Infect Dis. 2012; 31(11):3073-7. DOI: 10.1007/s10096-012-1667-x. View

3.
Beekmann S, Diekema D, Chapin K, Doern G . Effects of rapid detection of bloodstream infections on length of hospitalization and hospital charges. J Clin Microbiol. 2003; 41(7):3119-25. PMC: 165359. DOI: 10.1128/JCM.41.7.3119-3125.2003. View

4.
Mancini N, De Carolis E, Infurnari L, Vella A, Clementi N, Vaccaro L . Comparative evaluation of the Bruker Biotyper and Vitek MS matrix-assisted laser desorption ionization-time of flight (MALDI-TOF) mass spectrometry systems for identification of yeasts of medical importance. J Clin Microbiol. 2013; 51(7):2453-7. PMC: 3697679. DOI: 10.1128/JCM.00841-13. View

5.
van Veen S, Claas E, Kuijper E . High-throughput identification of bacteria and yeast by matrix-assisted laser desorption ionization-time of flight mass spectrometry in conventional medical microbiology laboratories. J Clin Microbiol. 2010; 48(3):900-7. PMC: 2832429. DOI: 10.1128/JCM.02071-09. View