Optimal Techniques for Magnetic Resonance Imaging of the Liver Using a Respiratory Navigator-gated Three-dimensional Spoiled Gradient-recalled Echo Sequence

Overview
Publisher Elsevier
Specialty Radiology
Date 2014 Jun 25
PMID 24960368
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To optimize the navigator-gating technique for the acquisition of high-quality three-dimensional spoiled gradient-recalled echo (3D SPGR) images of the liver during free breathing.

Materials And Methods: Ten healthy volunteers underwent 3D SPGR magnetic resonance imaging of the liver using a conventional navigator-gated 3D SPGR (cNAV-3D-SPGR) sequence or an enhanced navigator-gated 3D SPGR (eNAV-3D-SPGR) sequence. No exogenous contrast agent was used. A 20-ms wait period was inserted between the 3D SPGR acquisition component and navigator component of the eNAV-3D-SPGR sequence to allow T1 recovery. Visual evaluation and calculation of the signal-to-noise ratio were performed to compare image quality between the imaging techniques.

Result: The eNAV-3D-SPGR sequence provided better noise properties than the cNAV-3D-SPGR sequence visually and quantitatively. Navigator gating with an acceptance window of 2mm effectively inhibited respiratory motion artifacts. The widening of the window to 6mm shortened the acquisition time but increased motion artifacts, resulting in degradation of overall image quality. Neither slice tracking nor incorporation of short breath holding successfully compensated for the widening of the window.

Conclusion: The eNAV-3D-SPGR sequence with an acceptance window of 2mm provides high-quality 3D SPGR images of the liver.

Citing Articles

Correction of Arterial-Phase Motion Artifacts in Gadoxetic Acid-Enhanced Liver MRI Using an Innovative Unsupervised Network.

Pan F, Fan Q, Xie H, Bai C, Zhang Z, Chen H Bioengineering (Basel). 2023; 10(10).

PMID: 37892922 PMC: 10604307. DOI: 10.3390/bioengineering10101192.


DIfferential Subsampling With Cartesian Ordering With Respiratory Triggering Versus Conventional Liver Acquisition With Volume Acquisition: A Multiple Reader Preference Study.

Allen B, Ehieli W, Wildman-Tobriner B, Chaudhry M, Bozdogan E, Janas G J Comput Assist Tomogr. 2019; 43(4):623-627.

PMID: 31268878 PMC: 6629493. DOI: 10.1097/RCT.0000000000000888.


Pseudo-random Trajectory Scanning Suppresses Motion Artifacts on Gadoxetic Acid-enhanced Hepatobiliary-phase Magnetic Resonance Images.

Nakamura Y, Higaki T, Nishihara T, Harada K, Takizawa M, Bito Y Magn Reson Med Sci. 2019; 19(1):21-28.

PMID: 30880292 PMC: 7067909. DOI: 10.2463/mrms.mp.2018-0174.