Semi-adiabatic Shinnar-Le Roux Pulses and Their Application to Diffusion Tensor Imaging of Humans at 7T

Overview
Publisher Elsevier
Specialty Radiology
Date 2014 Jun 15
PMID 24928300
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

The adiabatic Shinnar-Le Roux (SLR) algorithm for radiofrequency (RF) pulse design enables systematic control of pulse parameters such as bandwidth, RF energy distribution and duration. Some applications, such as diffusion-weighted imaging (DWI) at high magnetic fields, would benefit from RF pulses that can provide greater B1 insensitivity while adhering to echo time and specific absorption rate (SAR) limits. In this study, the adiabatic SLR algorithm was employed to generate 6-ms and 4-ms 180° semi-adiabatic RF pulses which were used to replace the refocusing pulses in a twice-refocused spin echo (TRSE) diffusion-weighted echo planar imaging (DW-EPI) sequence to create two versions of a twice-refocused adiabatic spin echo (TRASE) sequence. The two versions were designed for different trade-offs between adiabaticity and echo time. Since a pair of identical refocusing pulses is applied, the quadratic phase imposed by the first is unwound by the second, preserving the linear phase created by the excitation pulse. In vivo images of the human brain obtained at 7Testa (7T) demonstrate that both versions of the TRASE sequence developed in this study achieve more homogeneous signal in the diffusion-weighted images than the conventional TRSE sequence. Semi-adiabatic SLR pulses offer a more B1-insensitive solution for diffusion preparation at 7T, while operating within SAR constraints. This method may be coupled with any EPI readout trajectory and parallel imaging scheme to provide more uniform coverage for diffusion tensor imaging at 7T and 3T.

Citing Articles

UltraHigh Field MR Imaging in Epilepsy.

Verma G, Delman B, Balchandani P Magn Reson Imaging Clin N Am. 2020; 29(1):41-52.

PMID: 33237014 PMC: 8794037. DOI: 10.1016/j.mric.2020.09.006.


Ultrahigh field MR Neuroimaging.

Verma G, Balchandani P Top Magn Reson Imaging. 2019; 28(3):137-144.

PMID: 31188272 PMC: 6784841. DOI: 10.1097/RMR.0000000000000210.


Ultrahigh field single-refocused diffusion weighted imaging using a matched-phase adiabatic spin echo (MASE).

Dyvorne H, OHalloran R, Balchandani P Magn Reson Med. 2015; 75(5):1949-57.

PMID: 26041650 PMC: 4670296. DOI: 10.1002/mrm.25790.


A SEmi-Adiabatic matched-phase spin echo (SEAMS) PINS pulse-pair for B1 -insensitive simultaneous multislice imaging.

Feldman R, Islam H, Xu J, Balchandani P Magn Reson Med. 2015; 75(2):709-17.

PMID: 25753055 PMC: 4979355. DOI: 10.1002/mrm.25654.


Ultra-High-Field MR Neuroimaging.

Balchandani P, Naidich T AJNR Am J Neuroradiol. 2014; 36(7):1204-15.

PMID: 25523591 PMC: 4472608. DOI: 10.3174/ajnr.A4180.

References
1.
Cox E, Gowland P . Simultaneous quantification of T2 and T'2 using a combined gradient echo-spin echo sequence at ultrahigh field. Magn Reson Med. 2010; 64(5):1440-5. DOI: 10.1002/mrm.22522. View

2.
Baum , Tycko , PINES . Broadband and adiabatic inversion of a two-level system by phase-modulated pulses. Phys Rev A Gen Phys. 1985; 32(6):3435-3447. DOI: 10.1103/physreva.32.3435. View

3.
Sigmund E, Gutman D . Diffusion-weighted imaging of the brain at 7 T with echo-planar and turbo spin echo sequences: preliminary results. Magn Reson Imaging. 2011; 29(6):752-65. DOI: 10.1016/j.mri.2011.02.016. View

4.
McNab J, Edlow B, Witzel T, Huang S, Bhat H, Heberlein K . The Human Connectome Project and beyond: initial applications of 300 mT/m gradients. Neuroimage. 2013; 80:234-45. PMC: 3812060. DOI: 10.1016/j.neuroimage.2013.05.074. View

5.
Setsompop K, Kimmlingen R, Eberlein E, Witzel T, Cohen-Adad J, McNab J . Pushing the limits of in vivo diffusion MRI for the Human Connectome Project. Neuroimage. 2013; 80:220-33. PMC: 3725309. DOI: 10.1016/j.neuroimage.2013.05.078. View