Background: International recommendations highlight the superior value of cardiac troponins (cTns) for early diagnosis of myocardial infarction along with analytical requirements of improved precision and detectability. In this multicenter study, we investigated the analytical performance of a new high sensitive cardiac troponin I (hs-cTnI) assay and its 99th percentile upper reference limit (URL).
Methods: Laboratories from nine European countries evaluated the ARCHITECT STAT high sensitive troponin I (hs-TnI) immunoassay on the ARCHITECT i2000SR/i1000SR immunoanalyzers. Imprecision, limit of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ) linearity of dilution, interferences, sample type, method comparisons, and 99th percentile URLs were evaluated in this study.
Results: Total imprecision of 3.3%-8.9%, 2.0%-3.5% and 1.5%-5.2% was determined for the low, medium and high controls, respectively. The lowest cTnI concentration corresponding to a total CV of 10% was 5.6 ng/L. Common interferences, sample dilution and carryover did not affect the hs-cTnI results. Slight, but statistically significant, differences with sample type were found. Concordance between the investigated hs-cTnI assay and contemporary cTnI assay at 99th percentile cut-off was found to be 95%. TnI was detectable in 75% and 57% of the apparently healthy population using the lower (1.1 ng/L) and upper (1.9 ng/L) limit of the LoD range provided by the ARCHITECT STAT hs-TnI package insert, respectively. The 99th percentile values were gender dependent.
Conclusions: The new ARCHITECT STAT hs-TnI assay with improved analytical features meets the criteria of high sensitive Tn test and will be a valuable diagnostic tool.
The impact of myocardial injury on outcomes in TAVI patients.
Stolte T, Lopez-Ayala P, Reichl J, Pfenniger A, Allegra G, Leibundgut G Clin Res Cardiol. 2024; .
PMID: 39663248 DOI: 10.1007/s00392-024-02585-1.
Lin Y, Zhan H, Wang S, Zhang G, Ye Z, Liu C iScience. 2024; 27(9):110643.
PMID: 39262795 PMC: 11388160. DOI: 10.1016/j.isci.2024.110643.
Urine high-sensitive troponin I in children cannot offer an applicable alternative to serum.
Bakos M, Dilber D, Jazbec A, Svagusa T, Potkonjak A, Braovac D Front Cardiovasc Med. 2024; 11:1391434.
PMID: 38836067 PMC: 11149416. DOI: 10.3389/fcvm.2024.1391434.
Albus M, Zimmermann T, Median D, Rumora K, Isayeva G, Amrein M Eur Heart J Cardiovasc Imaging. 2024; 25(9):1197-1205.
PMID: 38591997 PMC: 11346366. DOI: 10.1093/ehjci/jeae093.
Kloska S, Kozinski M, Stefanska A, Bergmann K, Mankowska-Cyl A, Siodmiak J J Med Biochem. 2023; 42(4):665-674.
PMID: 38084246 PMC: 10710818. DOI: 10.5937/jomb0-39081.