Performance Evaluation of the New Hematology Analyzer Sysmex XN-series

Overview
Specialty Hematology
Date 2014 May 13
PMID 24815300
Citations 22
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: The Sysmex XN-series is a new automated hematology analyzer designed to improve the accuracy of cell counts and the specificity of the flagging events.

Methods: The basic characteristics and the performance of new measurement channels of the XN were evaluated and compared with the Sysmex XE-2100 and the manual method. Fluorescent platelet count (PLT-F) was compared with the flow cytometric method. The low WBC mode and body fluid mode were also evaluated. For workflow analysis, 1005 samples were analyzed on both the XN and the XE-2100, and manual review rates were compared.

Results: All parameters measured by the XN correlated well with the XE-2100. PLT-F showed better correlation with the flow cytometric method (r(2)  = 0.80) compared with optical platelet count (r(2)  = 0.73) for platelet counts <70 × 10(9) /L. The low WBC mode reported accurate leukocyte differentials for samples with a WBC count <0.5 × 10(9) /L. Relatively good correlation was found for WBC counts between the manual method and the body fluid mode (r = 0.88). The XN made less flags than the XE-2100, while the sensitivities of both instruments were comparable.

Conclusion: The XN provided reliable results on low cell counts, as well as reduced manual blood film reviews, while maintaining a proper level of diagnostic sensitivity.

Citing Articles

Validation of low-volume blood collection tubes for routine hematologic testing.

Lippi G, Martini L, Cortivo B, Zecchetto C, Ferrari A J Med Biochem. 2025; 43(6):955-959.

PMID: 39876916 PMC: 11771976. DOI: 10.5937/jomb0-51438.


A comparative study of blood cell count in four automated hematology analyzers: An evaluation of the impact of preanalytical factors.

Astrand A, Wingren C, Walton C, Mattsson J, Agrawal K, Lindqvist M PLoS One. 2024; 19(5):e0301845.

PMID: 38787860 PMC: 11125483. DOI: 10.1371/journal.pone.0301845.


Evaluation of immature granulocyte parameters in myeloid neoplasms assayed by Sysmex XN hematology analyzer.

Lu Q, Li Y, Li T, Hou T, Zhao Y, Feng S J Hematop. 2024; 15(1):1-6.

PMID: 38358601 PMC: 10869398. DOI: 10.1007/s12308-022-00484-w.


Comparison of platelet count results on the Sysmex XN between citrate or MgSO and K2 EDTA anticoagulants.

Soulard M, Croix P, Cohen P Int J Lab Hematol. 2022; 45(1):20-28.

PMID: 36093831 PMC: 10087425. DOI: 10.1111/ijlh.13966.


Technological features of blast identification in the cerebrospinal fluid: A systematic review of flow cytometry and laboratory haematology methods.

Frater J, Shirai C, Brestoff J Int J Lab Hematol. 2022; 44 Suppl 1:45-53.

PMID: 35785436 PMC: 9463081. DOI: 10.1111/ijlh.13869.