» Articles » PMID: 24795621

Smaller, Softer, Lower-impedance Electrodes for Human Neuroprosthesis: a Pragmatic Approach

Overview
Journal Front Neuroeng
Specialty Neurology
Date 2014 May 6
PMID 24795621
Citations 19
Authors
Affiliations
Soon will be listed here.
Abstract

Finding the most appropriate technology for building electrodes to be used for long term implants in humans is a challenging issue. What are the most appropriate technologies? How could one achieve robustness, stability, compatibility, efficacy, and versatility, for both recording and stimulation? There are no easy answers to these questions as even the most fundamental and apparently obvious factors to be taken into account, such as the necessary mechanical, electrical and biological properties, and their interplay, are under debate. We present here our approach along three fundamental parallel pathways: we reduced electrode invasiveness and size without impairing signal-to-noise ratio, we increased electrode active surface area by depositing nanostructured materials, and we protected the brain from direct contact with the electrode without compromising performance. Altogether, these results converge toward high-resolution ECoG arrays that are soft and adaptable to cortical folds, and have been proven to provide high spatial and temporal resolution. This method provides a piece of work which, in our view, makes several steps ahead in bringing such novel devices into clinical settings, opening new avenues in diagnostics of brain diseases, and neuroprosthetic applications.

Citing Articles

Benchtop Performance of Novel Mixed Ionic-Electronic Conductive Electrode Form Factors for Biopotential Recordings.

Colachis M, Schlink B, Colachis 4th S, Shqau K, Huegen B, Palmer K Sensors (Basel). 2024; 24(10).

PMID: 38793990 PMC: 11125343. DOI: 10.3390/s24103136.


Soft and Conductive Polyethylene Glycol Hydrogel Electrodes for Electrocardiogram Monitoring.

Lee D, Song J, Kim J, Lee J, Son D, Shin M Gels. 2023; 9(12).

PMID: 38131943 PMC: 10742586. DOI: 10.3390/gels9120957.


Slippery Epidural ECoG Electrode for High-Performance Neural Recording and Interface.

Alahi M, Liu Y, Khademi S, Nag A, Wang H, Wu T Biosensors (Basel). 2022; 12(11).

PMID: 36421162 PMC: 9688081. DOI: 10.3390/bios12111044.


Biocompatibility Testing of Liquid Metal as an Interconnection Material for Flexible Implant Technology.

Foremny K, Nagels S, Kreienmeyer M, Doll T, Deferme W Nanomaterials (Basel). 2021; 11(12).

PMID: 34947600 PMC: 8706733. DOI: 10.3390/nano11123251.


Imaging the Efficiency of Poly(3,4-ethylenedioxythiophene) Doped with Acid-Functionalized Carbon Nanotube and Iridium Oxide Electrode Coatings for Microstimulation.

Zheng X, Yang Q, Vazquez A, Cui X Adv Nanobiomed Res. 2021; 1(7):2000092.

PMID: 34746928 PMC: 8552016. DOI: 10.1002/anbr.202000092.


References
1.
Rubehn B, Bosman C, Oostenveld R, Fries P, Stieglitz T . A MEMS-based flexible multichannel ECoG-electrode array. J Neural Eng. 2009; 6(3):036003. DOI: 10.1088/1741-2560/6/3/036003. View

2.
Quian Quiroga R, Nadasdy Z, Ben-Shaul Y . Unsupervised spike detection and sorting with wavelets and superparamagnetic clustering. Neural Comput. 2004; 16(8):1661-87. DOI: 10.1162/089976604774201631. View

3.
Gabay T, Ben-David M, Kalifa I, Sorkin R, Abrams Z, Ben-Jacob E . Electro-chemical and biological properties of carbon nanotube based multi-electrode arrays. Nanotechnology. 2009; 18(3):035201. DOI: 10.1088/0957-4484/18/3/035201. View

4.
Woerly S, Pinet E, de Robertis L, Bousmina M, Laroche G, Roitback T . Heterogeneous PHPMA hydrogels for tissue repair and axonal regeneration in the injured spinal cord. J Biomater Sci Polym Ed. 1998; 9(7):681-711. DOI: 10.1163/156856298x00091. View

5.
Erchova I, Lebedev M, Diamond M . Somatosensory cortical neuronal population activity across states of anaesthesia. Eur J Neurosci. 2002; 15(4):744-52. DOI: 10.1046/j.0953-816x.2002.01898.x. View