MRI-based Treatment Planning with Pseudo CT Generated Through Atlas Registration

Overview
Journal Med Phys
Specialty Biophysics
Date 2014 May 3
PMID 24784377
Citations 68
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To evaluate the feasibility and accuracy of magnetic resonance imaging (MRI)-based treatment planning using pseudo CTs generated through atlas registration.

Methods: A pseudo CT, providing electron density information for dose calculation, was generated by deforming atlas CT images previously acquired on other patients. The authors tested 4 schemes of synthesizing a pseudo CT from single or multiple deformed atlas images: use of a single arbitrarily selected atlas, arithmetic mean process using 6 atlases, and pattern recognition with Gaussian process (PRGP) using 6 or 12 atlases. The required deformation for atlas CT images was derived from a nonlinear registration of conjugated atlas MR images to that of the patient of interest. The contrasts of atlas MR images were adjusted by histogram matching to reduce the effect of different sets of acquisition parameters. For comparison, the authors also tested a simple scheme assigning the Hounsfield unit of water to the entire patient volume. All pseudo CT generating schemes were applied to 14 patients with common pediatric brain tumors. The image similarity of real patient-specific CT and pseudo CTs constructed by different schemes was compared. Differences in computation times were also calculated. The real CT in the treatment planning system was replaced with the pseudo CT, and the dose distribution was recalculated to determine the difference.

Results: The atlas approach generally performed better than assigning a bulk CT number to the entire patient volume. Comparing atlas-based schemes, those using multiple atlases outperformed the single atlas scheme. For multiple atlas schemes, the pseudo CTs were similar to the real CTs (correlation coefficient, 0.787-0.819). The calculated dose distribution was in close agreement with the original dose. Nearly the entire patient volume (98.3%-98.7%) satisfied the criteria of chi-evaluation (<2% maximum dose and 2 mm range). The dose to 95% of the volume and the percentage of volume receiving at least 95% of the prescription dose in the planning target volume differed from the original values by less than 2% of the prescription dose (root-mean-square, RMS < 1%). The PRGP scheme did not perform better than the arithmetic mean process with the same number of atlases. Increasing the number of atlases from 6 to 12 often resulted in improvements, but statistical significance was not always found.

Conclusions: MRI-based treatment planning with pseudo CTs generated through atlas registration is feasible for pediatric brain tumor patients. The doses calculated from pseudo CTs agreed well with those from real CTs, showing dosimetric accuracy within 2% for the PTV when multiple atlases were used. The arithmetic mean process may be a reasonable choice over PRGP for the synthesis scheme considering performance and computational costs.

Citing Articles

Clinical validation of MR-generated synthetic CT by MRCAT for brain tumor radiotherapy.

Yip T, Li Z, Li T J Appl Clin Med Phys. 2024; 26(1):e14494.

PMID: 39673148 PMC: 11713499. DOI: 10.1002/acm2.14494.


Model-based navigation of transcranial focused ultrasound neuromodulation in humans: Application to targeting the amygdala and thalamus.

Daneshzand M, Guerin B, Kotlarz P, Chou T, Dougherty D, Edlow B Brain Stimul. 2024; 17(4):958-969.

PMID: 39094682 PMC: 11367617. DOI: 10.1016/j.brs.2024.07.019.


MRI-only based material mass density and relative stopping power estimation via deep learning for proton therapy: a preliminary study.

Gao Y, Chang C, Mandava S, Marants R, Scholey J, Goette M Sci Rep. 2024; 14(1):11166.

PMID: 38750148 PMC: 11096170. DOI: 10.1038/s41598-024-61869-8.


Clinical feasibility of deep learning-based synthetic CT images from T2-weighted MR images for cervical cancer patients compared to MRCAT.

Kim H, Yoo S, Kim J, Kim Y, Lee J, Kim C Sci Rep. 2024; 14(1):8504.

PMID: 38605094 PMC: 11009270. DOI: 10.1038/s41598-024-59014-6.


A systematic literature review: deep learning techniques for synthetic medical image generation and their applications in radiotherapy.

Sherwani M, Gopalakrishnan S Front Radiol. 2024; 4:1385742.

PMID: 38601888 PMC: 11004271. DOI: 10.3389/fradi.2024.1385742.


References
1.
Schreibmann E, Nye J, Schuster D, Martin D, Votaw J, Fox T . MR-based attenuation correction for hybrid PET-MR brain imaging systems using deformable image registration. Med Phys. 2010; 37(5):2101-9. DOI: 10.1118/1.3377774. View

2.
Fallone B, Murray B, Rathee S, Stanescu T, Steciw S, Vidakovic S . First MR images obtained during megavoltage photon irradiation from a prototype integrated linac-MR system. Med Phys. 2009; 36(6):2084-8. DOI: 10.1118/1.3125662. View

3.
Hofmann M, Bezrukov I, Mantlik F, Aschoff P, Steinke F, Beyer T . MRI-based attenuation correction for whole-body PET/MRI: quantitative evaluation of segmentation- and atlas-based methods. J Nucl Med. 2011; 52(9):1392-9. DOI: 10.2967/jnumed.110.078949. View

4.
Chen L, Price Jr R, Wang L, Li J, Qin L, McNeeley S . MRI-based treatment planning for radiotherapy: dosimetric verification for prostate IMRT. Int J Radiat Oncol Biol Phys. 2004; 60(2):636-47. DOI: 10.1016/j.ijrobp.2004.05.068. View

5.
Stanescu T, Wachowicz K, Jaffray D . Characterization of tissue magnetic susceptibility-induced distortions for MRIgRT. Med Phys. 2012; 39(12):7185-93. DOI: 10.1118/1.4764481. View