» Articles » PMID: 24746788

Automatic Detection of Chest Compressions for the Assessment of CPR-quality Parameters

Overview
Journal Resuscitation
Specialty Emergency Medicine
Date 2014 Apr 22
PMID 24746788
Citations 16
Authors
Affiliations
Soon will be listed here.
Abstract

Aim: Accurate chest compression detection is key to evaluate cardiopulmonary resuscitation (CPR) quality. Two automatic compression detectors were developed, for the compression depth (CD), and for the thoracic impedance (TI). The objective was to evaluate their accuracy for compression detection and for CPR quality assessment.

Methods: Compressions were manually annotated using the force and ECG in 38 out-of-hospital resuscitation episodes, comprising 869 min and 67,402 compressions. Compressions were detected using a negative peak detector for the CD. For the TI, an adaptive peak detector based on the amplitude and duration of TI fluctuations was used. Chest compression rate (CC-rate) and chest compression fraction (CCF) were calculated for the episodes and for every minute within each episode. CC-rate for rescuer feedback was calculated every 8 consecutive compressions.

Results: The sensitivity and positive predictive value were 98.4% and 99.8% using CD, and 94.2% and 97.4% using TI. The mean CCF and CC-rate obtained from both detectors showed no significant differences with those obtained from the annotations (P>0.6). The Bland-Altman analysis showed acceptable 95% limits of agreement between the annotations and the detectors for the per-minute CCF, per-minute CC-rate, and CC-rate for feedback. For the detector based on TI, only 3.7% of CC-rate feedbacks had an error larger than 5%.

Conclusion: Automatic compression detectors based on the CD and TI signals are very accurate. In most cases, episode review could safely rely on these detectors without resorting to manual review. Automatic feedback on rate can be accurately done using the impedance channel.

Citing Articles

Registration of transthoracic impedance signal and ventilation volume data in out-of-hospital cardiac arrest.

Menant E, Tassart B, Meillier C, Lemoine F, Petermann A, Jost D Data Brief. 2025; 59:111345.

PMID: 39996051 PMC: 11848761. DOI: 10.1016/j.dib.2025.111345.


Re-arrest immediately after return of spontaneous circulation: A retrospective observational study of in-hospital cardiac arrest.

Unneland E, Norvik A, Bergum D, Buckler D, Bhardwaj A, Eftestol T Acta Anaesthesiol Scand. 2024; 69(1):e14567.

PMID: 39692065 PMC: 11653231. DOI: 10.1111/aas.14567.


Extracting physiologic and clinical data from defibrillators for research purposes to improve treatment for patients in cardiac arrest.

Nordseth T, Eftestol T, Aramendi E, Kvaloy J, Skogvoll E Resusc Plus. 2024; 18:100611.

PMID: 38524146 PMC: 10960142. DOI: 10.1016/j.resplu.2024.100611.


Impedance-Based Ventilation Detection and Signal Quality Control During Out-of-Hospital Cardiopulmonary Resuscitation.

Jaureguibeitia X, Aramendi E, Wang H, Idris A IEEE J Biomed Health Inform. 2023; 27(6):3026-3036.

PMID: 37028324 PMC: 10336723. DOI: 10.1109/JBHI.2023.3253780.


Quality of Cardiopulmonary Resuscitation and 5-Year Survival Following in-Hospital Cardiac Arrest.

Vestergaard L, Glerup Lauridsen K, Krarup N, Kristensen J, Andersen L, Lofgren B Open Access Emerg Med. 2021; 13:553-560.

PMID: 34938129 PMC: 8687881. DOI: 10.2147/OAEM.S341479.