Gradient Induced Artifacts in Simultaneous EEG-fMRI: Effect of Synchronization on Spiral and EPI K-space Trajectories

Overview
Publisher Elsevier
Specialty Radiology
Date 2014 Apr 22
PMID 24746775
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

The nature of the gradient induced electroencephalography (EEG) artifact is analyzed and compared for two functional magnetic resonance imaging (fMRI) pulse sequences with different k-space trajectories: echo planar imaging (EPI) and spiral. Furthermore, the performance of the average artifact subtraction algorithm (AAS) to remove the gradient artifact for both sequences is evaluated. The results show that the EEG gradient artifact for spiral sequences is one order of magnitude higher than for EPI sequences due to the chirping spectrum of the spiral sequence and the dB/dt of its crusher gradients. However, in the presence of accurate synchronization, the use of AAS yields the same artifact suppression efficiency for both pulse sequences below 80Hz. The quality of EEG signal after AAS is demonstrated for phantom and human data. EEG spectrogram and visual evoked potential (VEP) are compared outside the scanner and use both EPI and spiral pulse sequences. MR related artifact residues affect the spectra over 40Hz (less than 0.2 μV up to 120Hz) and modify the amplitude of P1, N2 and P300 in the VEP. These modifications in the EEG signal have to be taken into account when interpreting EEG data acquired in simultaneous EEG-fMRI experiments.

Citing Articles

Mindfulness Training Associated With Resting-State Electroencephalograms Dynamics in Novice Practitioners Mindful Breathing and Body-Scan.

Ng H, Wu C, Huang F, Cheng Y, Guu S, Huang C Front Psychol. 2021; 12:748584.

PMID: 34777144 PMC: 8581621. DOI: 10.3389/fpsyg.2021.748584.


Silent zero TE MR neuroimaging: Current state-of-the-art and future directions.

Ljungberg E, Damestani N, Wood T, Lythgoe D, Zelaya F, Williams S Prog Nucl Magn Reson Spectrosc. 2021; 123:73-93.

PMID: 34078538 PMC: 7616227. DOI: 10.1016/j.pnmrs.2021.03.002.


Exploring the relative efficacy of motion artefact correction techniques for EEG data acquired during simultaneous fMRI.

Daniel A, Smith J, Spencer G, Jorge J, Bowtell R, Mullinger K Hum Brain Mapp. 2018; 40(2):578-596.

PMID: 30339731 PMC: 6492138. DOI: 10.1002/hbm.24396.


Gradient Artefact Correction and Evaluation of the EEG Recorded Simultaneously with fMRI Data Using Optimised Moving-Average.

Ferreira J, Wu Y, Besseling R, Lamerichs R, Aarts R J Med Eng. 2016; 2016:9614323.

PMID: 27446943 PMC: 4942699. DOI: 10.1155/2016/9614323.


CXCR2 Inhibition Combined with Sorafenib Improved Antitumor and Antiangiogenic Response in Preclinical Models of Ovarian Cancer.

Devapatla B, Sharma A, Woo S PLoS One. 2015; 10(9):e0139237.

PMID: 26414070 PMC: 4587670. DOI: 10.1371/journal.pone.0139237.