» Articles » PMID: 24732635

Biodegradable Fe-based Alloys for Use in Osteosynthesis: Outcome of an in Vivo Study After 52 Weeks

Overview
Journal Acta Biomater
Publisher Elsevier
Date 2014 Apr 16
PMID 24732635
Citations 51
Authors
Affiliations
Soon will be listed here.
Abstract

This study investigates the degradation performance of three Fe-based materials in a growing rat skeleton over a period of 1 year. Pins of pure Fe and two Fe-based alloys (Fe-10 Mn-1Pd and Fe-21 Mn-0.7C-1Pd, in wt.%) were implanted transcortically into the femur of 38 Sprague-Dawley rats and inspected after 4, 12, 24 and 52 weeks. The assessment was performed by ex vivo microfocus computed tomography, weight-loss determination, surface analysis of the explanted pins and histological examination. The materials investigated showed signs of degradation; however, the degradation proceeded rather slowly and no significant differences between the materials were detected. We discuss these unexpected findings on the basis of fundamental considerations regarding iron corrosion. Dense layers of degradation products were formed on the implants' surfaces, and act as barriers against oxygen transport. For the degradation of iron, however, the presence of oxygen is an indispensable prerequisite. Its availability is generally a critical factor in bony tissue and rather limited there, i.e. in the vicinity of our implants. Because of the relatively slow degradation of both pure Fe and the Fe-based alloys, their suitability for bulk temporary implants such as those in osteosynthesis applications appears questionable.

Citing Articles

On the addition of Au and Pt to a Fe-Mn-Si alloy for biodegradable implants.

Lemke J, Fiocchi J, Biffi C, Coda A, Tuissi A Heliyon. 2025; 11(4):e42663.

PMID: 40066028 PMC: 11891682. DOI: 10.1016/j.heliyon.2025.e42663.


Degradation behavior of austenite, ferrite, and martensite present in biodegradable Fe-based alloys in three protein-rich pseudo-physiological solutions.

Cherqaoui A, Cao Q, Gatto M, Paternoster C, Mengucci P, Mantovani D Bioact Mater. 2024; 41:96-107.

PMID: 39734418 PMC: 11681225. DOI: 10.1016/j.bioactmat.2024.06.025.


Study of the effect of magnetic fields on static degradation of Fe and Fe-12Mn-1.2C in balanced salts modified Hanks' solution.

Limon I, Multigner M, Paternoster C, Lieblich M, Torres B, Mantovani D Bioact Mater. 2024; 40:524-540.

PMID: 39040567 PMC: 11261407. DOI: 10.1016/j.bioactmat.2024.06.027.


Biodegradable high-nitrogen iron alloy anastomotic staples: and studies.

Lu S, Wang P, Wang Q, Deng P, Yuan Y, Fu X Bioact Mater. 2024; 40:34-46.

PMID: 38910967 PMC: 11190430. DOI: 10.1016/j.bioactmat.2024.06.005.


Patient-specific implants made of 3D printed bioresorbable polymers at the point-of-care: material, technology, and scope of surgical application.

Maintz M, Tourbier C, de Wild M, Cattin P, Beyer M, Seiler D 3D Print Med. 2024; 10(1):13.

PMID: 38639834 PMC: 11031859. DOI: 10.1186/s41205-024-00207-0.