Validation of Anatomically Shaped Cranial Collimation (ACC) in Orthodontic Lateral Cephalography

Overview
Date 2014 Apr 12
PMID 24720607
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

The use of an anatomically shaped cranial collimator (ACC) to reduce patient dose in orthodontic lateral cephalography was investigated in this study. The aim was to evaluate the potential interference of the ACC on landmark identification for orthodontic cephalometry. Consecutive orthodontic patients underwent a total of 100 cephalograms using an ACC mounted on a Veraviewepocs(®) 3D X550 (J. Morita Co., Kyoto, Japan) X-ray unit. 10 observers were asked whether the identification of 5 landmarks close to the collimated area was hindered or rendered impossible by the presence of the collimator. Of the 500 landmarks that were judged by the 10 observers, 496 (99.2%) were reported to lack hindrance. In three landmarks, a minority of the observers reported hindrance. In 1 landmark, 8 of the 10 observers reported hindrance by the collimator. In no instance did the observers state that the identification of landmarks was impossible as a result of the collimation. Application of the ACC on the cephalostat of the X-ray unit is a viable way of reducing patient dose, as it only marginally interferes with the diagnostic yield of the exposure. The need to retake images when the ACC is applied was found to be extremely low.

Citing Articles

Dose reduction in orthodontic lateral cephalography: dosimetric evaluation of a novel cephalographic thyroid protector (CTP) and anatomical cranial collimation (ACC).

Hoogeveen R, Rottke D, van der Stelt P, Berkhout W Dentomaxillofac Radiol. 2015; 44(4):20140260.

PMID: 25564885 PMC: 4628428. DOI: 10.1259/dmfr.20140260.

References
1.
LAbee E, Tan H . Improved radiation hygiene in lateral cephalometry and a method to obtain good reproduction of the soft tissue profile. J Clin Orthod. 1982; 16(6):381-6. View

2.
Hujoel P, Hollender L, Bollen A, Young J, Cunha-Cruz J, McGee M . Thyroid shields and neck exposures in cephalometric radiography. BMC Med Imaging. 2006; 6:6. PMC: 1543616. DOI: 10.1186/1471-2342-6-6. View

3.
Gijbels F, Sanderink G, Wyatt J, van Dam J, Nowak B, Jacobs R . Radiation doses of collimated vs non-collimated cephalometric exposures. Dentomaxillofac Radiol. 2003; 32(2):128-33. DOI: 10.1259/dmfr/33233723. View

4.
Mandall N, OBrien K, Worthington H . Radiation reduction using a modified collimated lateral skull radiograph during orthodontic treatment. Clin Orthod Res. 2000; 2(4):179-85. DOI: 10.1111/ocr.1999.2.4.179. View

5.
Alcaraz M, Garcia-Vera M, Bravo L, Martinez-Beneyto Y, Armero D, Morant J . Collimator with filtration compensator: clinical adaptation to meet European Union recommendation 4F on radiological protection for dental radiography. Dentomaxillofac Radiol. 2009; 38(6):413-20. DOI: 10.1259/dmfr/15580890. View