» Articles » PMID: 24676383

Multispectral Optoacoustic Tomography at 64, 128, and 256 Channels

Overview
Journal J Biomed Opt
Date 2014 Mar 29
PMID 24676383
Citations 41
Authors
Affiliations
Soon will be listed here.
Abstract

Optoacoustic (photoacoustic) imaging has already showcased the capacity to offer high-resolution small animal visualization in vivo in a variety of cancer, cardiovascular, or neuroimaging applications. In particular, multispectral optoacoustic tomography (MSOT) has shown imaging along the spectral and the time dimensions, enabling sensing of multiple molecules over time and, more recently, in real time. Furthermore, cross-sectional imaging of at least 20 mm diameter has been showcased in vivo in animals and humans using 64-element curved transducers placed along a single curved line. Herein, we investigated the imaging improvements gained by utilizing a larger number of detectors and inquired whether more detectors will result in measurable image quality improvements. For this reason, we implemented MSOT using 64-, 128-, and 256-element transducers and imaged the same phantoms and animals under similar conditions. Further, corroborated by numerical simulation analysis, our findings quantify the improvements in resolution and overall image quality for the increasing number of detectors used pointing to significant improvements in image quality for the 256 detector array, over 64 or 128 detectors.

Citing Articles

Guided Multispectral Optoacoustic Tomography for 3D Imaging of the Murine Colon.

Buehler A, Brown E, Eckstein M, Thoma O, Wachter F, Mandelbaum H Adv Sci (Weinh). 2025; 12(10):e2413434.

PMID: 39836529 PMC: 11905093. DOI: 10.1002/advs.202413434.


Recent Advances in Photoacoustic Imaging: Current Status and Future Perspectives.

Liu H, Teng X, Yu S, Yang W, Kong T, Liu T Micromachines (Basel). 2024; 15(8).

PMID: 39203658 PMC: 11356134. DOI: 10.3390/mi15081007.


Image reconstruction of multispectral sparse sampling photoacoustic tomography based on deep algorithm unrolling.

Ge J, Mo Z, Zhang S, Zhang X, Zhong Y, Liang Z Photoacoustics. 2024; 38:100618.

PMID: 38957484 PMC: 11217744. DOI: 10.1016/j.pacs.2024.100618.


Photoacoustic imaging for cutaneous melanoma assessment: a comprehensive review.

Fakhoury J, Benavides Lara J, Manwar R, Zafar M, Xu Q, Engel R J Biomed Opt. 2024; 29(Suppl 1):S11518.

PMID: 38223680 PMC: 10785699. DOI: 10.1117/1.JBO.29.S1.S11518.


Imaging increased metabolism in the spinal cord in mice after middle cerebral artery occlusion.

Ni R, Straumann N, Fazio S, Dean-Ben X, Louloudis G, Keller C Photoacoustics. 2023; 32:100532.