Evaluation of an Automated Microplate Technique in the Galileo System for ABO and Rh(D) Blood Grouping
Authors
Affiliations
Background: A number of automated devices for pretransfusion testing have recently become available. This study evaluated the Immucor Galileo System, a fully automated device based on the microplate hemagglutination technique for ABO/Rh (D) determinations.
Methods: Routine ABO/Rh typing tests were performed on 13,045 samples using the Immucor automated instruments. Manual tube method was used to resolve ABO forward and reverse grouping discrepancies. D-negative test results were investigated and confirmed manually by the indirect antiglobulin test (IAT).
Results: The system rejected 70 tests for sample inadequacy. 87 samples were read as "No-type-determined" due to forward and reverse grouping discrepancies. 25 tests gave these results because of sample hemolysis. After further tests, we found 34 tests were caused by weakened RBC antibodies, 5 tests were attributable to weak A and/or B antigens, 4 tests were due to mixed-field reactions, and 8 tests had high titer cold agglutinin with blood qualifications which react only at temperatures below 34 degrees C. In the remaining 11 cases, irregular RBC antibodies were identified in 9 samples (seven anti-M and two anti-P) and two subgroups were identified in 2 samples (one A1 and one A2) by a reference laboratory. As for D typing, 2 weak D+ samples missed by automated systems gave negative results, but weak-positive reactions were observed in the IAT.
Conclusions: The Immucor Galileo System is reliable and suited for ABO and D blood groups, some reasons may cause a discrepancy in ABO/D typing using a fully automated system. It is suggested that standardization of sample collection may improve the performance of the fully automated system.
Microfluidic chip designs and their application for E antigen typing on red blood cells.
Maraming P, Shean Aye N, Panyakakaew P, Tippayawat P, Daduang S, Choowongkomon K RSC Adv. 2025; 15(8):6077-6088.
PMID: 39995458 PMC: 11848836. DOI: 10.1039/d4ra08321k.
Higuchi M, Sekiba Y, Watanabe N Clin Hemorheol Microcirc. 2023; 84(1):33-41.
PMID: 37005880 PMC: 10357158. DOI: 10.3233/CH-221651.
Li H, Guo K Front Med (Lausanne). 2022; 9:827619.
PMID: 35223922 PMC: 8873177. DOI: 10.3389/fmed.2022.827619.
ABO Mistyping of -AB Blood Group by the Automated Microplate Technique.
Chun S, Ryu M, Cha S, Seo J, Cho D Transfus Med Hemother. 2018; 45(1):5-10.
PMID: 29593455 PMC: 5836188. DOI: 10.1159/000475506.