» Articles » PMID: 24656798

Clinical Results of an EPID-based In-vivo Dosimetry Method for Pelvic Cancers Treated by Intensity-modulated Radiation Therapy

Overview
Journal Phys Med
Date 2014 Mar 25
PMID 24656798
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

The purpose of our work was to investigate the feasibility of using an EPID-based in-vivo dosimetry method initially designed for conformal fields on pelvic dynamic IMRT fields. The method enables a point dose delivered to the patient to be calculated from the transit signal acquired with an electronic portal imaging device (EPID). After defining a set of correction factors allowing EPID pixel values to be converted into absolute doses, several tests on homogeneous water-equivalent phantoms were performed to estimate the validity of the method in reference conditions. The effects of different treatment parameters, such as delivered dose, field size dependence and patient thickness were also studied. The model was first evaluated on a group of 53 patients treated by 3D conformal radiotherapy (3DCRT) and then on 92 patients treated by IMRT, both for pelvic cancers. For each measurement, the dose was reconstructed at the isocenter (DREC) and compared with the dose calculated by our treatment planning system (DTPS). Excellent agreement was found between DREC and DTPS for both techniques. For 3DCRT treatments, the mean deviation between DREC and DTPS for the 211 in-vivo dose verifications was equal to -1.0  ±  2.2% (1SD). Concerning IMRT treatments, the averaged deviation for the 418 fields verified was equal to -0.3 ± 2.6% (1SD) proving that the method is able to reconstruct a dose for dynamic IMRT pelvic fields. Based on these results, tolerance criteria and action levels were established before its implementation in clinical routine.

Citing Articles

Analyses of integrated EPID images for on-treatment quality assurance to account for interfractional variations in volumetric modulated arc therapy.

Matsushita N, Nakamura M, Sasaki M, Yano S, Yoshimura M, Mizowaki T J Appl Clin Med Phys. 2020; 21(1):110-116.

PMID: 31909889 PMC: 6964755. DOI: 10.1002/acm2.12805.


Epid-based in vivo dose verification for lung stereotactic treatments delivered with multiple breath-hold segmented volumetric modulated arc therapy.

Cilla S, Ianiro A, Craus M, Viola P, Deodato F, Macchia G J Appl Clin Med Phys. 2019; 20(3):37-44.

PMID: 30790439 PMC: 6414179. DOI: 10.1002/acm2.12538.


A Feasibility Study for in vivo Dosimetry Procedure in Routine Clinical Practice.

Falco M, Giancaterino S, De Nicola A, Adorante N, Gimenez De Lorenzo R, DI Tommaso M Technol Cancer Res Treat. 2018; 17:1533033818779201.

PMID: 29871570 PMC: 5992805. DOI: 10.1177/1533033818779201.


On-line estimations of delivered radiation doses in three-dimensional conformal radiotherapy treatments of carcinoma uterine cervix patients in linear accelerator.

Putha S, Saxena P, Banerjee S, Srinivas C, Vadhiraja B, Ravichandran R J Med Phys. 2017; 41(4):224-233.

PMID: 28144114 PMC: 5228045. DOI: 10.4103/0971-6203.195186.


A dual two dimensional electronic portal imaging device transit dosimetry model based on an empirical quadratic formalism.

Tan Y, Metwaly M, Glegg M, Baggarley S, Elliott A Br J Radiol. 2015; 88(1051):20140645.