Performance Evaluation of Platelet Counting by Novel Fluorescent Dye Staining in the XN-series Automated Hematology Analyzers

Overview
Journal J Clin Lab Anal
Publisher Wiley
Date 2014 Mar 21
PMID 24648166
Citations 12
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Conventional automated hematology analyzers have limitations in platelet measurements such as poor accuracy and precision in the low count range and interference by nonplatelet particles. In order to improve it, the newly developed XN-Series automated hematology analyzers (Sysmex Corporation, Kobe, Japan) have been installed with a new dedicated channel for platelet analysis (PLT-F), which is based on a fluorescence flow cytometry method with uses of a novel fluorescent dye specifically staining platelets. We evaluated the basic performance of this new PLT-F channel.

Methods: Basic performance of the PLT-F channel in within-run reproducibility and assay linearity was studied using standard methods. Correlation was studied between PLT-F and a conventional automated hematology analyzer (XE-2100) and immunoplatelet analysis using anti-CD61 monoclonal antibody (Cell-Dyn Sapphire; Abbott Laboratories). The assay interference by nonplatelet particles such as fragmented red and white blood cells was evaluated by using clinical samples, respectively, from burn injury and acute leukemia.

Results: Basic performance of the PLT-F platelet counting was satisfactory in within-run reproducibility, linearity and correlation with the conventional analyzer. The correlation was satisfactory also with the immunoplatelet analysis, even for samples from a patient with burn injury, and those with white blood cell fragments displayed, platelet abnormal flag and low platelet counts (<50 × 10(9)/l).

Conclusion: The platelet counting performance of the PLT-F channel of the XN Series had improved accuracy and precision in the low range and in abnormal samples, avoiding the interference by nonplatelet particles.

Citing Articles

Reference guide for the diagnosis of adult primary immune thrombocytopenia, 2023 edition.

Kashiwagi H, Kuwana M, Murata M, Shimada N, Takafuta T, Yamanouchi J Int J Hematol. 2023; 119(1):1-13.

PMID: 37957517 PMC: 10770234. DOI: 10.1007/s12185-023-03672-1.


Stability over time of immature platelet fraction and comparison between EDTA and citrated whole blood samples.

Noel M, Feugray G, Kasonga F, Buchonnet G, Daliphard S, Barbay V J Clin Lab Anal. 2023; 37(13-14):e24946.

PMID: 37475131 PMC: 10492457. DOI: 10.1002/jcla.24946.


Comparison of platelet count results on the Sysmex XN between citrate or MgSO and K2 EDTA anticoagulants.

Soulard M, Croix P, Cohen P Int J Lab Hematol. 2022; 45(1):20-28.

PMID: 36093831 PMC: 10087425. DOI: 10.1111/ijlh.13966.


Alinity hq platelet count is not impacted by severe microcytosis.

Lakos G, Mukhtar Z, Masi L, Valente S, Papa F J Clin Lab Anal. 2022; 36(4):e24218.

PMID: 35274768 PMC: 8993633. DOI: 10.1002/jcla.24218.


Evaluation of automated hematology analyzer DYMIND DH76 compared to SYSMEX XN 1000 system.

Velizarova M, Yacheva T, Genova M, Svinarov D J Med Biochem. 2021; 40(4):367-377.

PMID: 34616226 PMC: 8451233. DOI: 10.5937/jomb0-28836.


References
1.
Segal H, Briggs C, Kunka S, Casbard A, Harrison P, Machin S . Accuracy of platelet counting haematology analysers in severe thrombocytopenia and potential impact on platelet transfusion. Br J Haematol. 2005; 128(4):520-5. DOI: 10.1111/j.1365-2141.2004.05352.x. View

2.
Harrison P, Ault K, Chapman S, Charie L, Davis B, Fujimoto K . An interlaboratory study of a candidate reference method for platelet counting. Am J Clin Pathol. 2001; 115(3):448-59. DOI: 10.1309/91PR-E4G6-XBAF-N8DY. View

3.
Grimaldi E, Del Vecchio L, Scopacasa F, Pardo C, Capone F, Pariante S . Evaluation of the platelet counting by Abbott CELL-DYN SAPPHIRE haematology analyser compared with flow cytometry. Int J Lab Hematol. 2008; 31(2):151-60. DOI: 10.1111/j.1751-553X.2007.01015.x. View

4.
Briggs C, Longair I, Kumar P, Singh D, Machin S . Performance evaluation of the Sysmex haematology XN modular system. J Clin Pathol. 2012; 65(11):1024-30. DOI: 10.1136/jclinpath-2012-200930. View

5.
Li S, Salhany K . Spurious elevation of automated platelet counts in secondary acute monocytic leukemia associated with tumor lysis syndrome. Arch Pathol Lab Med. 1999; 123(11):1111-4. DOI: 10.5858/1999-123-1111-SEOAPC. View