In Vivo Left-ventricular Contact Force Analysis: Comparison of Antegrade Transseptal with Retrograde Transaortic Mapping Strategies and Correlation of Impedance and Electrical Amplitude with Contact Force

Abstract

Aims: Clinical outcomes following radiofrequency ablation of ventricular tachycardias (VTs) depend on catheter tip-to-tissue contact force (CF). Left-ventricular (LV) mapping is performed via antegrade-transseptal or retrograde-transaortic approaches, and the applied CF may depend on the approach used. This study evaluated (i) the impact of antegrade-transseptal vs. retrograde-transaortic LV-mapping approaches on CF and catheter stability and (ii) the clinical value of the commonly used surrogate markers of catheter-myocardial contact-impedance, unipolar, and bipolar electrogram amplitudes.

Methods And Results: An antegrade-transseptal and a retrograde-transaortic LV-mapping approach was performed in 10 patients undergoing VT ablation by using CF-sensing catheters. Operators were blinded to CF data and data were analysed according to 11 predefined LV segments. Three thousand three hundred and twenty-four mapping points (1577 antegrade, 1747 retrograde) were analysed, including 80 (2.4%) points with maximum CF > 100 g. Median antegrade and retrograde CF were 16.0 g (q1-q3; 8.4-26.2) and 15.3 g (9.8-23.4), respectively. Contact force was significantly higher antegradely in mid-anteroseptum, mid-lateral, and apical segments, and significantly higher retrogradely in basal-anteroseptum, basal-inferoseptum, basal-inferior, and basal-lateral segments. Contact force did correlate with impedance, unipolar, and bipolar electrogram amplitudes; however, there were large overlaps.

Conclusions: Antegrade vs. retrograde LV-mapping approaches result in different CF. A combined approach to the LV mapping may improve the overall LV mapping, potentially resulting in better clinical outcomes for the left VT catheter ablation. The previous surrogate markers used to assess CF do correlate with in vivo CF; however, due to a larger overlap, their clinical value is limited.

Citing Articles

Best practices in robotic magnetic navigation-guided catheter ablation of cardiac arrhythmias, a position paper of the Society for Cardiac Robotic Navigation.

Noten A, Szili-Torok T, Ernst S, Burkhardt D, Cavaco D, Chen X Front Cardiovasc Med. 2024; 11:1431396.

PMID: 39399515 PMC: 11466809. DOI: 10.3389/fcvm.2024.1431396.


Is catheter-tissue contact force value important for ablation of ventricular arrhythmias originating from the left ventricular papillary muscles?.

Baran J, Skrzynska-Kowalczyk M, Piotrowski R, Sikorska A, Krynski T, Kulakowski P Front Cardiovasc Med. 2023; 10:1166810.

PMID: 37273878 PMC: 10235700. DOI: 10.3389/fcvm.2023.1166810.


[Catheter ablation of ventricular tachycardias in patients with ischemic cardiomyopathy].

Bourier F, Busch S, Sommer P, Maurer T, Althoff T, Shin D Herzschrittmacherther Elektrophysiol. 2022; 33(1):88-97.

PMID: 35157112 DOI: 10.1007/s00399-022-00845-z.


Updates in Ventricular Tachycardia Ablation.

Campbell T, Bennett R, Kotake Y, Kumar S Korean Circ J. 2020; 51(1):15-42.

PMID: 33377327 PMC: 7779814. DOI: 10.4070/kcj.2020.0436.


Multipolar mapping for catheter ablation of premature ventricular complexes originating from papillary muscles in the structurally normal heart: a case series.

Cai C, Wang J, Niu H, Hua W, Chu J, Zhang S BMC Cardiovasc Disord. 2020; 20(1):464.

PMID: 33115432 PMC: 7594480. DOI: 10.1186/s12872-020-01747-z.


References
1.
Kumar S, Morton J, Halloran K, Spence S, Lee G, Wong M . Effect of respiration on catheter-tissue contact force during ablation of atrial arrhythmias. Heart Rhythm. 2012; 9(7):1041-1047.e1. DOI: 10.1016/j.hrthm.2012.02.015. View

2.
Kumar S, Morton J, Lee J, Halloran K, Spence S, Gorelik A . Prospective characterization of catheter-tissue contact force at different anatomic sites during antral pulmonary vein isolation. Circ Arrhythm Electrophysiol. 2012; 5(6):1124-9. DOI: 10.1161/CIRCEP.112.972208. View

3.
Reddy V, Neuzil P, Taborsky M, Ruskin J . Short-term results of substrate mapping and radiofrequency ablation of ischemic ventricular tachycardia using a saline-irrigated catheter. J Am Coll Cardiol. 2003; 41(12):2228-36. DOI: 10.1016/s0735-1097(03)00492-3. View

4.
Shah D, Lambert H, Langenkamp A, Vanenkov Y, Leo G, Gentil-Baron P . Catheter tip force required for mechanical perforation of porcine cardiac chambers. Europace. 2010; 13(2):277-83. DOI: 10.1093/europace/euq403. View

5.
Berruezo A, Mont L, Nava S, Chueca E, Bartholomay E, Brugada J . Electrocardiographic recognition of the epicardial origin of ventricular tachycardias. Circulation. 2004; 109(15):1842-7. DOI: 10.1161/01.CIR.0000125525.04081.4B. View