» Articles » PMID: 24484064

Radiation Dose Index of Renal Colic Protocol CT Studies in the United States: a Report from the American College of Radiology National Radiology Data Registry

Overview
Journal Radiology
Specialty Radiology
Date 2014 Feb 4
PMID 24484064
Citations 29
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To determine radiation dose indexes for computed tomography (CT) performed with renal colic protocols in the United States, including frequency of reduced-dose technique usage and any institutional-level factors associated with high or low dose indexes.

Materials And Methods: The Dose Imaging Registry (DIR) collects deidentified CT data, including examination type and dose indexes, for CT performed at participating institutions; thus, the DIR portion of the study was exempt from institutional review board approval and was HIPAA compliant. CT dose indexes were examined at the institutional level for CT performed with a renal colic protocol at institutions that contributed at least 10 studies to the registry as of January 2013. Additionally, patients undergoing CT for renal colic at a single institution (with institutional review board approval and informed consent from prospective subjects and waiver of consent from retrospective subjects) were studied to examine individual renal colic CT dose index patterns and explore relationships between patient habitus, demographics, and dose indexes. Descriptive statistics were used to analyze dose indexes, and linear regression and Spearman correlations were used to examine relationships between dose indexes and institutional factors.

Results: There were 49 903 renal colic protocol CT examinations conducted at 93 institutions between May 2011 and January 2013. Mean age ± standard deviation was 49 years ± 18, and 53.9% of patients were female. Institutions contributed a median of 268 (interquartile range, 77-699) CT studies. Overall mean institutional dose-length product (DLP) was 746 mGy ⋅ cm (effective dose, 11.2 mSv), with a range of 307-1497 mGy ⋅ cm (effective dose, 4.6-22.5 mSv) for mean DLPs. Only 2% of studies were conducted with a DLP of 200 mGy ⋅ cm or lower (a "reduced dose") (effective dose, 3 mSv), and only 10% of institutions kept DLP at 400 mGy ⋅ cm (effective dose, 6 mSv) or less in at least 50% of patients.

Conclusion: Reduced-dose renal protocol CT is used infrequently in the United States. Mean dose index is higher than reported previously, and institutional variation is substantial.

Citing Articles

Dual-energy CT kidney stone characterization-can diagnostic accuracy be achieved at low radiation dose?.

Euler A, Wullschleger S, Sartoretti T, Muller D, Keller E, Lavrek D Eur Radiol. 2023; 33(9):6238-6244.

PMID: 36988716 PMC: 10415460. DOI: 10.1007/s00330-023-09569-1.


Feasibility of photon-counting CT for femoroacetabular impingement syndrome evaluation: lower radiation dose and improved diagnostic confidence.

Ferrero A, Powell G, Adaaquah D, Rajendran K, Thorne J, Krych A Skeletal Radiol. 2023; 52(9):1651-1659.

PMID: 36971838 DOI: 10.1007/s00256-023-04325-4.


Computed Tomography Radiation Exposure among Urinary Tract Stone Patients at Tikur Anbessa Specialized Hospital: A Retrospective Study.

Dellie S, Nebiyou T, Tadesse D, Kebede T Ethiop J Health Sci. 2022; 32(Spec Iss 1):33-40.

PMID: 36339957 PMC: 9624098. DOI: 10.4314/ejhs.v32i1.6S.


Application of deep learning reconstruction of ultra-low-dose abdominal CT in the diagnosis of renal calculi.

Zhang X, Zhang G, Xu L, Bai X, Zhang J, Xu M Insights Imaging. 2022; 13(1):163.

PMID: 36209195 PMC: 9547757. DOI: 10.1186/s13244-022-01300-w.


Diagnostic reference levels for chest computed tomography in children as a function of patient size.

Bos D, Zensen S, Opitz M, Haubold J, Nassenstein K, Kinner S Pediatr Radiol. 2022; 52(8):1446-1455.

PMID: 35378606 PMC: 9271112. DOI: 10.1007/s00247-022-05340-8.


References
1.
Ljunghall S, Danielson B . A prospective study of renal stone recurrences. Br J Urol. 1984; 56(2):122-4. DOI: 10.1111/j.1464-410x.1984.tb05346.x. View

2.
Meagher T, Sukumar V, Collingwood J, Crawley T, Schofield D, Henson J . Low dose computed tomography in suspected acute renal colic. Clin Radiol. 2001; 56(11):873-6. DOI: 10.1053/crad.2001.0842. View

3.
Liu W, Esler S, Kenny B, Goh R, Rainbow A, Stevenson G . Low-dose nonenhanced helical CT of renal colic: assessment of ureteric stone detection and measurement of effective dose equivalent. Radiology. 2001; 215(1):51-4. DOI: 10.1148/radiology.215.1.r00ap4051. View

4.
Brenner D, Hall E . Computed tomography--an increasing source of radiation exposure. N Engl J Med. 2007; 357(22):2277-84. DOI: 10.1056/NEJMra072149. View

5.
Morin R, Coombs L, Chatfield M . ACR Dose Index Registry. J Am Coll Radiol. 2011; 8(4):288-91. DOI: 10.1016/j.jacr.2010.12.022. View