» Articles » PMID: 24418729

The Vexing Problem of Thrombosis in Long-term Mechanical Circulatory Support

Overview
Publisher Elsevier
Date 2014 Jan 15
PMID 24418729
Citations 40
Authors
Affiliations
Soon will be listed here.
Abstract

Durable left ventricular assist devices (LVADs) have not only enhanced longevity but also conferred sustained improvements in quality of life, symptom control, and functional capacity in patients with medically refractory advanced heart failure. Problems with device-related infection, bleeding, neurologic events, right-sided heart failure, and device malfunction have dominated the clinical care of patients living on mechanical support. Even as adoption of durable LVADs accelerated globally, we began to encounter a growing dilemma of pump malfunction caused by thrombosis. In early 2011, clinicians began to notice a spike in the incidence of pump thrombosis with the HeartMate II (Thoratec Corp, Pleasanton, CA) LVAD. By 2012, the problem of thrombosis in LVADs began to consume most of the scientific direction as centers and collaborative groups began to dissect this nascent phenomenon. In this perspective, we describe the magnitude and implications of pump thrombosis, discuss secular and management trends in this unique population, attempt to dissect the problem at its root, offer guidance on surveillance and therapeutic principles, and outline issues that deserve our immediate and collaborative attention.

Citing Articles

A Computational Investigation of the Effects of Temporal Synchronization of Left Ventricular Assist Device Speed Modulation with the Cardiac Cycle on Intraventricular Hemodynamics.

Straccia A, Chassagne F, Barbour M, Beckman J, Li S, Mahr C Ann Biomed Eng. 2024; 52(6):1763-1778.

PMID: 38517620 DOI: 10.1007/s10439-024-03489-x.


Depression among Patients with an Implanted Left Ventricular Assist Device: Uncovering Pathophysiological Mechanisms and Implications for Patient Care.

Alnsasra H, Khalil F, Perue R, Azab A Int J Mol Sci. 2023; 24(14).

PMID: 37511030 PMC: 10379142. DOI: 10.3390/ijms241411270.


In Vitro Investigation of the Effect of the Timing of Left Ventricular Assist Device Speed Modulation on Intraventricular Flow Patterns.

Chassagne F, Beckman J, Li S, Mahr C, Aliseda A ASAIO J. 2023; 69(6):533-543.

PMID: 36881637 PMC: 11187697. DOI: 10.1097/MAT.0000000000001893.


Evolution of thrombolytic therapy in patients with HeartWare ventricular assist device thrombosis: a single-institutional experience.

Fox H, Gyoten T, Rojas S, Lauenroth V, Gunther S, Schramm R Interact Cardiovasc Thorac Surg. 2022; 35(1).

PMID: 35234899 PMC: 9252132. DOI: 10.1093/icvts/ivac054.


A Multi-Domain Simulation Study of a Pulsatile-Flow Pump Device for Heart Failure With Preserved Ejection Fraction.

Ozturk C, Rosalia L, Roche E Front Physiol. 2022; 13:815787.

PMID: 35145432 PMC: 8822361. DOI: 10.3389/fphys.2022.815787.