EPID in Vivo Dosimetry in RapidArc Technique

Overview
Specialty Oncology
Date 2013 Dec 31
PMID 24376916
Citations 12
Authors
Affiliations
Soon will be listed here.
Abstract

Aim: The aim of the study was to estimate the dose at the reference point applying an aSi-EPID device in the course of patient treatment.

Materials And Methods: The method assumes direct proportionality between EPID signal and dose delivered to the patient reference point during the treatment session. The procedure consists of treatment plan calculation for the actual patient in the arc technique. The plan was realized with an elliptic water-equivalent phantom. An ionization chamber inside the phantom measured the dose delivered to the reference point. Simultaneously, the EPID matrix measured the CU distribution. EPID signal was also registered during patient irradiation with the same treatment plan. The formula for in vivo dose calculation was based on the CU(g) function, EPID signal registered during therapy and the relation between the dose and EPID signal level measured for the phantom. In vivo dose was compared with dose planned with the treatment planning system. Irradiation was performed with a Clinac accelerator by Varian Medical Systems in the RapidArc technique. The Clinac was equipped with an EPID matrix (electronic portal image device) of aSi-1000. Treatment plans were calculated with the Eclipse/Helios system. The phantom was a Scanditronix/Wellhöfer Slab phantom, and the ionization chamber was a 0.6 ccm PTW chamber.

Results: In vivo dose calculations were performed for five patients. Planned dose at the reference point was 2 Gy for each treatment plan. Mean in vivo dose was in the range of 1.96-2.09.

Conclusions: Our method was shown to be appropriate for in vivo dose evaluation in the RapidArc technique.

Citing Articles

Portal dosimetry in radiotherapy repeatability evaluation.

Slosarek K, Plaza D, Nas A, Reudelsdorf M, Wendykier J, Bekman B J Appl Clin Med Phys. 2020; 22(1):156-164.

PMID: 33314643 PMC: 7856497. DOI: 10.1002/acm2.13123.


Two-dimensional solid-state array detectors: A technique for in vivo dose verification in a variable effective area.

Utitsarn K, Biasi G, Stansook N, Alrowaili Z, Petasecca M, Carolan M J Appl Clin Med Phys. 2019; 20(11):88-94.

PMID: 31609090 PMC: 6839376. DOI: 10.1002/acm2.12744.


EPID-based daily verification of reproducibility of patients' irradiation with IMRT plans.

Kruszyna-Mochalska M Rep Pract Oncol Radiother. 2018; 23(5):309-314.

PMID: 30108458 PMC: 6088814. DOI: 10.1016/j.rpor.2018.05.003.


Diode-based transmission detector for IMRT delivery monitoring: a validation study.

Li T, Wu Q, Matzen T, Yin F, ODaniel J J Appl Clin Med Phys. 2016; 17(5):235-244.

PMID: 27685115 PMC: 5874094. DOI: 10.1120/jacmp.v17i5.6204.


Clinical examples of 3D dose distribution reconstruction, based on the actual MLC leaves movement, for dynamic treatment techniques.

Osewski W, Dolla L, Radwan M, Szlag M, Rutkowski R, Smolinska B Rep Pract Oncol Radiother. 2014; 19(6):420-7.

PMID: 25337416 PMC: 4201771. DOI: 10.1016/j.rpor.2014.04.013.


References
1.
Fogliata A, Clivio A, Nicolini G, Vanetti E, Cozzi L . Intensity modulation with photons for benign intracranial tumours: a planning comparison of volumetric single arc, helical arc and fixed gantry techniques. Radiother Oncol. 2008; 89(3):254-62. DOI: 10.1016/j.radonc.2008.07.021. View

2.
Otto K . Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys. 2008; 35(1):310-7. DOI: 10.1118/1.2818738. View

3.
Lagerwaard F, Meijer O, van der Hoorn E, Verbakel W, Slotman B, Senan S . Volumetric modulated arc radiotherapy for vestibular schwannomas. Int J Radiat Oncol Biol Phys. 2009; 74(2):610-5. DOI: 10.1016/j.ijrobp.2008.12.076. View

4.
Verbakel W, Cuijpers J, Hoffmans D, Bieker M, Slotman B, Senan S . Volumetric intensity-modulated arc therapy vs. conventional IMRT in head-and-neck cancer: a comparative planning and dosimetric study. Int J Radiat Oncol Biol Phys. 2009; 74(1):252-9. DOI: 10.1016/j.ijrobp.2008.12.033. View

5.
Nicolini G, Vanetti E, Clivio A, Fogliata A, Korreman S, Bocanek J . The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy. Radiat Oncol. 2008; 3:24. PMC: 2553075. DOI: 10.1186/1748-717X-3-24. View