Radial Artery Applanation Tonometry for Continuous Non-invasive Arterial Pressure Monitoring in Intensive Care Unit Patients: Comparison with Invasively Assessed Radial Arterial Pressure

Overview
Journal Br J Anaesth
Publisher Elsevier
Specialty Anesthesiology
Date 2013 Dec 21
PMID 24355832
Citations 22
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Radial artery applanation tonometry technology can be used for continuous non-invasive measurement of arterial pressure (AP). The purpose of this study was to evaluate this AP monitoring technology in intensive care unit (ICU) patients in comparison with invasive AP monitoring using a radial arterial catheter.

Methods: In 24 ICU patients (German university hospital), AP values were simultaneously recorded on a beat-to-beat basis using radial artery applanation tonometry (T-Line system; Tensys Medical, San Diego, CA, USA) and a radial arterial catheter (contralateral arm). The primary endpoint of the study was to investigate the accuracy and precision of the non-invasively assessed AP measurements with the Bland-Altman method based on averaged 10 beat AP epochs (n=2993 10 beat epochs).

Results: For mean AP (MAP), systolic AP (SAP), and diastolic AP (DAP), we observed a bias (±standard deviation of the bias; 95% limits of agreement; percentage error) of +2 mm Hg (±6; -11 to +15 mm Hg; 15%), -3 mm Hg (±15; -33 to +27 mm Hg; 23%), and +5 mm Hg (±7; -9 to +19 mm Hg; 22%), respectively.

Conclusions: In ICU patients, MAP and DAP measurements obtained using radial artery applanation tonometry show clinically acceptable agreement with invasive AP determination with a radial arterial catheter. While the radial artery applanation tonometry technology also allows SAP measurements with high accuracy, its precision for SAP measurements needs to be further improved.

Citing Articles

Contemporary Review of Hemodynamic Monitoring in the Critical Care Setting.

Rali A, Butcher A, Tedford R, Sinha S, Mekki P, Van Spall H US Cardiol. 2024; 16:e12.

PMID: 39600839 PMC: 11588176. DOI: 10.15420/usc.2021.34.


Patent landscape review of non-invasive medical sensors for continuous monitoring of blood pressure and their validation in critical care practice.

Litvinova O, Bilir A, Parvanov E, Niebauer J, Kletecka-Pulker M, Kimberger O Front Med (Lausanne). 2023; 10:1138051.

PMID: 37497278 PMC: 10366595. DOI: 10.3389/fmed.2023.1138051.


Perioperative Validation of the TensorTip™ MTX Device for Noninvasive Arterial Pressure Measurement: A Method Comparison Study.

Servaas S, van Eijk L, de Vreede S, Malagon I, Slagt C Cardiol Res. 2023; 13(6):372-379.

PMID: 36660063 PMC: 9822676. DOI: 10.14740/cr1438.


Noninvasive continuous arterial pressure monitoring during anesthesia induction in patients undergoing cardiac surgery.

Frank P, Logemann F, Gras C, Palmaers T Ann Card Anaesth. 2021; 24(3):281-287.

PMID: 34269255 PMC: 8404578. DOI: 10.4103/aca.ACA_120_20.


Novel wearable and contactless monitoring devices to identify deteriorating patients in the clinical setting: a systematic review protocol.

Chan P, McNeil J, Nguyen T, Ryan N, Hopper I Syst Rev. 2020; 9(1):104.

PMID: 32375867 PMC: 7201723. DOI: 10.1186/s13643-020-01370-1.