Reproducibility of Small Animal Cine and Scar Cardiac Magnetic Resonance Imaging Using a Clinical 3.0 Tesla System

Overview
Journal BMC Med Imaging
Publisher Biomed Central
Specialty Radiology
Date 2013 Dec 19
PMID 24345214
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

Background: To evaluate the inter-study, inter-reader and intra-reader reproducibility of cardiac cine and scar imaging in rats using a clinical 3.0 Tesla magnetic resonance (MR) system.

Methods: Thirty-three adult rats (Sprague-Dawley) were imaged 24 hours after surgical occlusion of the left anterior descending coronary artery using a 3.0 Tesla clinical MR scanner (Philips Healthcare, Best, The Netherlands) equipped with a dedicated 70 mm solenoid receive-only coil. Left-ventricular (LV) volumes, mass, ejection fraction and amount of myocardial scar tissue were measured. Intra-and inter-observer reproducibility was assessed in all animals. In addition, repeat MR exams were performed in 6 randomly chosen rats within 24 hours to assess inter-study reproducibility.

Results: The MR imaging protocol was successfully completed in 32 (97%) animals. Bland-Altman analysis demonstrated high intra-reader reproducibility (mean bias%: LV end-diastolic volume (LVEDV), -1.7%; LV end-systolic volume (LVESV), -2.2%; LV ejection fraction (LVEF), 1.0%; LV mass, -2.7%; and scar mass, -1.2%) and high inter-reader reproducibility (mean bias%: LVEDV, 3.3%; LVESV, 6.2%; LVEF, -4.8%; LV mass, -1.9%; and scar mass, -1.8%). In addition, a high inter-study reproducibility was found (mean bias%: LVEDV, 0.1%; LVESV, -1.8%; LVEF, 1.0%; LV mass, -4.6%; and scar mass, -6.2%).

Conclusions: Cardiac MR imaging of rats yielded highly reproducible measurements of cardiac volumes/function and myocardial infarct size on a clinical 3.0 Tesla MR scanner system. Consequently, more widely available high field clinical MR scanners can be employed for small animal imaging of the heart e.g. when aiming at serial assessments during therapeutic intervention studies.

Citing Articles

Evaluation of Cardiac Function in Young Mdx Mice Using MRI with Feature Tracking and Self-Gated Magnetic Resonance Cine Imaging.

Ueda J, Saito S Diagnostics (Basel). 2023; 13(8).

PMID: 37189573 PMC: 10137836. DOI: 10.3390/diagnostics13081472.


Application of Magnetic Resonance Strain Analysis Using Feature Tracking in a Myocardial Infarction Model.

Onishi R, Ueda J, Ide S, Koseki M, Sakata Y, Saito S Tomography. 2023; 9(2):871-882.

PMID: 37104142 PMC: 10141923. DOI: 10.3390/tomography9020071.

References
1.
Yang Z, Berr S, Gilson W, Toufektsian M, French B . Simultaneous evaluation of infarct size and cardiac function in intact mice by contrast-enhanced cardiac magnetic resonance imaging reveals contractile dysfunction in noninfarcted regions early after myocardial infarction. Circulation. 2004; 109(9):1161-7. DOI: 10.1161/01.CIR.0000118495.88442.32. View

2.
Dietrich T, Hucko T, Bourayou R, Jahnke C, Paetsch I, Atrott K . High resolution magnetic resonance imaging in atherosclerotic mice treated with ezetimibe. Int J Cardiovasc Imaging. 2009; 25(8):827-36. DOI: 10.1007/s10554-009-9487-5. View

3.
Fieno D, Jaffe W, Simonetti O, Judd R, Finn J . TrueFISP: assessment of accuracy for measurement of left ventricular mass in an animal model. J Magn Reson Imaging. 2002; 15(5):526-31. DOI: 10.1002/jmri.10107. View

4.
Rehr R, Malloy C, Filipchuk N, Peshock R . Left ventricular volumes measured by MR imaging. Radiology. 1985; 156(3):717-9. DOI: 10.1148/radiology.156.3.4023232. View

5.
Kim R, Chen E, Lima J, Judd R . Myocardial Gd-DTPA kinetics determine MRI contrast enhancement and reflect the extent and severity of myocardial injury after acute reperfused infarction. Circulation. 1996; 94(12):3318-26. DOI: 10.1161/01.cir.94.12.3318. View